A.T.S. 3000 AUTOMATIC TOURNIQUET SYSTEM
19
Table 2.3 Error Codes
ERROR CODE
CRITICAL FAILURE
EXPLANATION/APPROPRIATE ACTIONS
E006
SYSTEM FAIL
A Safety Monitor detection error was detected. Cycle the ON/
STANDBY button. If problem persists, note error code and contact
manufacturer.
E007
MATH FAIL
A processor error was detected. Cycle the ON/STANDBY button. If
problem persists, note error code and contact manufacturer.
E008
SYSTEM FAIL
The processor received an unknown interrupt. Cycle the
ON/STANDBY button. If problem persists, note error code and
contact manufacturer.
E010
ROM FAIL
A non-volatile memory value was detected during power-up.
Cycle the ON/STANDBY button. If problem persists, note error
code and contact manufacturer.
E011
ROM FAIL
ROM check error detected a self-test. Cycle the ON/STANDBY
button. If problem persists, note error code and contact manufacturer.
E012
RAM FAIL
A microprocessor RAM error was detected during self-test. Cycle
the ON/STANDBY button. If problem persists, note error code and
contact manufacturer.
E013
BATTERY FAIL
Battery voltage below minimum threshold. Plug unit in and cycle
the ON/STANDBY button. If problem persists, note error code and
contact manufacturer.
E014
BATTERY FAIL
A battery charging circuit failure was detected at power-up. Verify a
battery is connected and cycle the ON/STANDBY button. If problem
persists, note error code and contact manufacturer.
E015
CAL FAIL
Incorrect zero pressure calibration point was detected. Calibrate unit.
If problem persists, note error code and contact manufacturer.
E016
CAL FAIL
Incorrect calibration point was detected. Calibrate unit. If problem
persists, note error code and contact manufacturer.
E017
CAL FAIL
Calibration error was detected. Calibrate unit. If problem persists, note
error code and contact manufacturer.
E018
CAL M FAIL
Main cuff transducer calibration error was detected. Calibrate unit. If
problem persists, note error code and contact manufacturer.
E019
CAL 2 FAIL
Second cuff transducer calibration error was detected. Calibrate unit.
If problem persists, note error code and contact manufacturer.
E020 XXXX1
SYSTEM FAIL
A software exception error was detected. Cycle the ON/STANDBY
button. If problem persists, note error code and contact manufacturer.
E021 XXXX1
AMP FAIL
A transducer reference voltage is out of range. Calibrate unit. If
problem persists, note error code and contact manufacturer.
E022 XXXX1
AMP FAIL
Reservoir transducer is out of range. Calibrate unit. If problem
persists, note error code and contact manufacturer.
E023 XXXX1
AMP FAIL
The Main cuff transducer is out of range. Calibrate unit. If problem
persists, note error code and contact manufacturer.
E024 XXXX1
AMP FAIL
The Second cuff transducer is out of range. Calibrate unit. If problem
persists, note error code and contact manufacturer.
E025 XXXX1
AMP FAIL
A voltage level is out of range. Cycle the ON/STANDBY button. If
problem persists, note error code and contact manufacturer.
E026 XXXX1
AMP FAIL
A voltage level is out of range. Cycle the ON/STANDBY button. If
problem persists, note error code and contact manufacturer.
E027 XXXX1
SYSTEM FAIL
A speaker or alarm tone generation circuitry failure was detected.
Verify that the speaker is connected and cycle the ON/STANDBY
button. If problem persists, note error code and contact manufacturer.
Continued