Table 5.6:
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The pods are intended for use in the electromagnetic environment specified below.
The customer or the user of the pods should assure that it is used in such an environment.
Immunity test
IEC 60601test level
Compliance level
Electromagnetic
environment – guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV
air
± 2 kV, ± 4 kV, ± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
Floors should be wood,
concrete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for input power port
(AC and DC)
100 kHz repetition frequency
± 1 kV for signal input/output
parts
100 kHz repetition frequency
Not applicable
Not applicable
Surge IEC
61000-4-5
± 0,5 kV, ± 1 kV line-to-line for
input power port (AC and DC)
± 0,5 kV, ± 1 kV, ± 2 kV line-to-
ground for input power port
(AC and DC)
± 2 kV line-to-ground for
signal input/output parts
Not applicable
Not applicable
Voltage dips on power supply
lines IEC 61000-4-11
0 % U_T; 0,5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0 % U_T; 1 cycle
and
70 % U_T; 25/30 cycles
Single phase: at 0°
Not applicable
Not applicable
Voltage interruptions on power
supply lines IEC 61000-4-11
0 % U_T; 250/300 cycles
Not applicable
Not applicable
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
50 Hz and 60 Hz
(because it is battery
powered)
30 A/m
50 Hz and 60 Hz
(because it is battery powered)
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
NOTE: UT is the AC mains voltage prior to application of the test level.