4 |
iASSIST 2-Pod Version
Surgical Technique
Figure 2.2
System Controller Computer Setup
System Controller Computer Setup
1. Open and place the system controller carrying
case on a stable flat surface outside the sterile
zone at a maximum distance of four meters from
the instruments and OR tables.
Warning:
The system controller computer is
not intended to be used in the sterile zone; and
must be connected to an AC outlet during the
entire procedure.
The system controller computer must not be
placed in a sterile bag as it might overheat.
2. Secure the inside flap in place to maintain the
appropriate angle for the system controller
computer.
3. Position the system controller computer on the
angled flap.
4. Connect the power cord to the power supply and
computer port. Plug it in to an AC outlet.
5. Press the power button (red button) located at
the top left hand corner of the system controller
computer to turn on the computer.
6. The home screen will be displayed after boot-up is
completed.