D2O FL1 Product Manual - 17
5.8 DISCONNECTING FROM A COMPUTER
Once all programming and commissioning processes have been completed it
is important to disconnect the
4-Pin M12 to USB-A Cable
and replace the
Dust
Cap
to improve the environmental durability of the product and protect the
connector.
The
4-Pin M12 to USB-A Cable
should be kept and stored safely for future use.
IMPORTANT NOTE:
Ensure the D2O’s
Battery Power
has been enabled through
the CLI before leaving the D2O for its normal operation. No
samples or transmissions will occur if the
Battery Power
is
off.
If the unit is being preconfigured prior to its final installation it may be
desirable to preserve the battery life until deployment. The D2O’s Battery
Power can be disabled by entering the below command. This is an
Admin
Restricted c
ommand and cannot be set until the
admin
command has been
executed:
sys -p OFF
Note: the unit will remain powered by the USB connection until disconnected.
There are many additional CLI commands that can be reviewed by entering
help
into the CLI, or by reviewing the Serial Commands section of this
manual.
It may be desirable to generate a Commissioning Report once installed to
verify the system settings are as intended. This can be generated using the
report
command. This will generate a list of currently set parameters which
can be copied and saved for record keeping purposes.
IMPORTANT NOTE:
Ensure the D2O’s
Battery Power
has been enabled through
the CLI before leaving the D2O for its normal operation. No
samples or transmissions will occur if the
Battery Power
is
off.