background image

32

33

7. 

DATA LOGGING

7.7.3 CURRENT CALIBRATION RECORDS

Each calibration
record contains 
the following 
(See Figure 9):

• Date and time

that calibration
took place

• Pre-calibration

readings (for up 
to 4 gases)

• Post-calibration readings

(for up to 4 gases)

• An asterisk symbol (*) to indicate a note is attached

Note: For calibration readings, “Fail” indicates a calibration failure, and “N/A” indicates that

a gas sensor is missing.

• To display the current calibration records, click on the View menu and then on Current

Calibration (or simply click on the View Current Calibration button on the toolbar of the
Main Screen). A list of all calibration records is displayed for the instrument in that
download (See Figure 9).

• To delete a calibration record, select the record and click on the Delete Record button.

• To add a note to a calibration record, select the record and click on the Add/View Note

button. It brings up the Calibration Note dialog box. In this dialog box, type the note and
then click on Save. An asterisk (*) is displayed at the end of the record, indicating that a
note is attached.

Note: An ellipse mark (…) at the end of a record indicates that a note can be associated
with that record.

• To print the current calibration records for that download while on the Current

Calibration dialog box, check on the “w/ Note” check box to indicate all the attached
notes are to be printed along with the records (or uncheck it to print without the notes)
and click on Print Records.

7. 

DATA LOGGING

7.7.4 CALIBRATION HISTORY

Each calibration record
contains the following
(See Figure 10):

• Date and time for

which calibration
takes place

• Pre-calibration

readings (for 
up to 4 gases)

• Post-calibration

readings (for 
up to 4 gases)

• An asterisk symbol 

(*) to indicate
a note is attached

Note: For calibration

readings, “Fail”
indicates a calibration 
failure, and “N/A” indicates 
that a gas sensor is missing.

• To display the calibration history, click on the View menu and then on Calibration History

(or simply click on the Calibration History button on the toolbar of the Main Screen). 
A calibration history is displayed for the selected instrument (See Figure 10).

• To select an instrument, select the serial number from the Serial Number drop-down 

box, or enter the serial number manually in the box and click Find. For example, for an
instrument with serial number 1234, append two 0’s in the front to make it a 6-digit
number, i.e. 001234, and click Find.

• To delete a calibration record, select the record and click on the Delete Record button.

• To add a note to a calibration record, select the record and click on the Add/View Note

button. It brings up the Calibration Note dialog box. In this dialog box, type the note and
then click on Save. An asterisk (*) is displayed at the end of the record, indicating that a
note is attached.

Note

An ellipse mark (…) at the end of a record indicates that a note can be associated with

that record.

• To print the calibration history while on the Calibration History dialog box, check on the

“w/ Note” check box to indicate all the attached notes are to be printed along with the
history (or uncheck it to print without the notes) and click on Print History.

M U L T I   G A S D E T E C T I O N

WARNING 

Keep in mind that current calibration records are subsets of the calibration history, so any

records deleted from either the Current Calibration dialog or the Calibration History dialog

are permanently removed.

WARNING 

Keep in mind that the current calibration records are subsets of the calibration history, so

any records deleted from either the Current Calibration dialog or the Calibration History

dialog are permanently removed.

Figure 10

A screenshot of the Calibration History dialog box.

Figure 9 

A screenshot of the Current Calibration dialog box. 

Summary of Contents for Lumidor MicroMAX+

Page 1: ...M U LT I G A S D E T E C T I O N U S E R A N D M A I N T E N A N C E M A N U A L...

Page 2: ...ty claim is valid Zellweger Analytics will repair or replace the defective product free of charge and send it or any replacement back The decision to repair or replace parts shall be determined by Zel...

Page 3: ...6 4 Auto zeroing 24 TABLE OF CONTENTS SUBJECT SECTION PAGE 6 5 Selecting Pump Mode Diffusion Mode 24 6 6 Selecting Immediate Alarm Mode TWA Alarm Mode 24 6 7 Restoring Factory Defaults 25 6 7 Calibra...

Page 4: ...ems Packaging retain for possible future use Instrument with internal rechargeable battery pack Calibration certificate Standard accessories AC battery charger Alkaline battery holder Dust filter 10 f...

Page 5: ...more than an hour after which time a low battery alarm is indicated by two beeps every 15 seconds Low battery alarm will continue for at least half an hour after which instrument will beep continuous...

Page 6: ...eak reading highest toxic or combustible and lowest oxygen levels STEL 15 minute accumulated short term exposure level and TWA 8 hour time weighted average since instrument was last turned on The STEL...

Page 7: ...bypass each setting depress the MODE button To change any setting depress the ON OFF button For example to enter calibration mode depress the ON OFF button when CALIBRATE is displayed See Section 5 Ca...

Page 8: ...to instrument Place hose in area to be sampled 6 Wait two minutes for full response 7 Take readings 8 Retrieve hose and disconnect from instrument Hose lengths up to 100 feet may be used for pretesti...

Page 9: ...sulfides or silicone containing lubricants present in the atmosphere being monitored The safest possible course of action is to expose the sensors to a known concentration test gas before each day s...

Page 10: ...lay CALIBRATING and count down to 0 7 When CAL COMPLETE is displayed observe which sensors are calibrated and displayed below CAL COMPLETE Remove gas and turn off gas flow 8 Depress the ON OFF switch...

Page 11: ...the menu The following are all built in programmable options Date format date time daylight savings time Auto zero during start up sequence TWA alarm on or off Factory default settings Calibration Pu...

Page 12: ...value If the date format is MM DD the first two digits on the left represent the current month and the first digit must be either a 0 or a 1 Let us set the date to September 2 2002 We need 09 to repre...

Page 13: ...efault values for these alarms are pre set See Appendix A Specifications Section I User Programmable High and Low Alarm Set Points at the factory but are user programmable 2 In the TWA alarm mode thre...

Page 14: ...7 DATA LOGGING 7 4 UNINSTALL PROCEDURE For Windows XP users go to Start Control Panel For other Windows users go to Start Settings Control Panel When the Control Panel opens double click the Add Remo...

Page 15: ...first name and location is 14 14 and 20 alphanumeric characters respectively Downloading at the same date and time for a particular instrument is not allowed For example if you download data from an i...

Page 16: ...LOGGING 7 7 2 GRAPH SCREEN Information on the Graph Screen includes the following See Figure 8 User information employee s name instrument s serial number and user ID Download information date time an...

Page 17: ...on readings Fail indicates a calibration failure and N A indicates that a gas sensor is missing To display the calibration history click on the View menu and then on Calibration History or simply clic...

Page 18: ...ption to OFF The factory alarm setting is immediate alarm with values chosen to provide earlier warning than the STEL and TWA settings The immediate high and immediate low alarms are user programmable...

Page 19: ...TING LOCATION The Edit Location function allows you to modify the location for a specific download After a data file is opened click on the Edit menu then on Edit Location or simply click on the Edit...

Page 20: ...rams such as Microsoft Excel Figure 12 A screenshot of the Import File dialog box Figure 13 A screenshot of the Archive Data dialog box 7 DATA LOGGING 7 13 RECOVERING DATA The Recover Data function en...

Page 21: ...tached to the sensor pins 7 Align pins of new sensor with sockets and push down to secure sensor 8 Reconnect gas plate and secure with screw 9 Reconnect battery connector 10 Install case cover and sec...

Page 22: ...bing Part number GFV194 8 9 SAMPLING HOSE MAINTENANCE Check water trap Part number GFV196 the spherical disc located near the middle of the sampling hose periodically to make sure it is clean and unob...

Page 23: ...nd is the only potentiometer trimpot in the instrument 4 Set digital voltmeter to the DC voltage range black lead to test point 12 analog ground and the red lead to test point 10 These test points are...

Page 24: ...ble 90db at 1 foot Visual Large LED bar Flashing display characters APPENDIX A SPECIFICATIONS G GAS RANGES AND RESOLUTION SENSOR INSTRUMENT RANGE RESOLUTION Combustible LEL 0 100 LEL 1 LEL Oxygen O2 0...

Page 25: ...PPM Hydrogen Sulfide H2S 20PPM 15PPM 10PPM Sulfur Dioxide SO2 10 0PPM 5 0PPM 2 0PPM Phosphine PH3 5 0PPM 1 0PPM 0 3PPM Ammonia NH3 50PPM 35PPM 25PPM Chlorine CL2 5 0PPM 1 0PPM 0 5PPM Chlorine Dioxide...

Page 26: ...10 000 1 Ethylene 500 15 Carbon Monoxide 100 7 Carbon Dioxide 10 000 3 Hydrogen Sulfide 25 25 APPENDIX A SPECIFICATIONS SULPHUR DIOXIDE GAS CONCENTRATION PPM TYPICAL RESPONSE PPM Carbon Monoxide 300pp...

Page 27: ...ermane 1 0 Hydrogen Fluoride 4 0 Alcohols 1000 0 NOTE 1 Short term in minute range APPENDIX A SPECIFICATIONS HYDROGEN CYANIDE GAS CONCENTRATION PPM TYPICAL RESPONSE PPM Hydrogen Sulfide 10 40 Sulfur D...

Page 28: ...Hydrogen Fluoride 3 0 Alcohols 1000 0 Ozone 0 1 0 03 Note Short gas exposure in minute range APPENDIX A SPECIFICATIONS CHLORINE GAS CONCENTRATION PPM TYPICAL RESPONSE PPM Hydrogen Sulfide 10 Neg 0 3 N...

Page 29: ...carbons 10 000 0 Ammonia 30 0 Ozone 1 0 7 Diborane 0 1 0 Arsine 0 1 0 Germane 1 0 Fluorine 3 0 Alcohols 1000 0 Nitrogen Monoxide 100 0 COMBO TOX CO H2S GAS CONCENTRATION PPM TYPICAL RESPONSE PPM Hydro...

Page 30: ...rounding the instrument Refer to Section 8 8 Pump Inlet Filter Replacement for maintenance DO NOT USE non standard sampling tube or fittings Storage Temperature Range 20 degrees C to 50 degrees C 4 de...

Page 31: ...0 P11 52 B Note Max time from Turn On to Ready does not exceed 80 seconds APPENDIX C MODE SEQUENCE STANDARD SEQUENCE DISPLAY LED SPEAKER TO CLEAR PEAK VALUES TO ENTER USER SETUP BEEP 3 BEEPS BEEP BEEP...

Page 32: ...Sequence Standard Mode for how to enter User Set up Mode A DISPLAY LED SPEAKER DEFAULT ON BEEP PUMP STATUS ON APPENDIX D MODE SEQUENCE USER SET UP MODE CALIBRATION LED SPEAKER BEEP 3 BEEPS BEEP OX LEL...

Page 33: ...25 PURGING PLEASE WAIT OX LEL C0 H2S APPLY GAS AUTOMATIC MEMORY CLEAR AFTER PURGING CALIBRATION COMPLETE ONLY SHOWS GASES CALIBRATED NORMAL OPERATING MODE ON OFF YES NO B APPENDIX D MODE SEQUENCE USER...

Page 34: ...P 08 49 B B B FLASH CONTINUOUS HORN FLASH LONG CONTINUOUS BEEPS FLASH LOW GAS ALARM LOW FLOW ALARM 3 BEEPS OX LEL C0 H2S 20 9 0 0 0 DANGER P 08 49 B 3 BEEPS FLASH HIGH GAS ALARM APPENDIX E ALARMS MOD...

Page 35: ...carrying case MCC215A MPRO KIT 1 MINI Kit containing 34 liter calibration gas cylinder GFV243 CH4 CO H2S balance air regulator tubing and water resistant carrying case MCC210A MPRO KIT 2 Kit containi...

Page 36: ...d by the foot GFV256 Tygon hose tubing sold by the foot GFV196 Replacement water filter APPENDIX I REPLACEMENT PARTS PART NO DESCRIPTION 20HFC Replacement dust filter MPLUS DATALOG Data logging softwa...

Page 37: ...first calibration gas is GFV123 the second calibration gas is either GFV 213 58 Liters or GFV 250 34 Liters b MPLUS 4ABCH O2 LEL CO and SO2 Follow the procedure in Section 5 2 Calibration Procedure w...

Page 38: ...M U LT I G A S D E T E C T I O N 9021 Zellweger Analytics 400 Sawgrass Corporate Pkwy Suite 100 Sunrise FL 33325 Tel 1 954 514 2700 Toll free 1 800 538 0363 Fax 1 954 514 2784 MAN MPLUS REV A 0...

Reviews: