Functionality

Troubleshooting the UVClean

A flashing red LED indicates a device fault and the ZEISS 
UVClean will not run until the fault is cleared.  

Please refer to the steps below to clear possible fault 
conditions.

Troubleshooting Steps

Step 1 – Reset Device

Turn off power switch on the back of the ZEISS UVClean and 
wait 15 seconds. Turn the device back on. Close lid and press 
the Start button on top of the lid to begin a disinfection cycle.  

If the LED indicator continues to flash red, proceed to Step 2.

Step 2 – Reseat UVC Bulbs

Turn off the ZEISS UVClean and unplug from the electrical 
outlet. Open the lid and locate the four UVC bulbs inside the 
device. Wearing the gloves provided, hold the two white end 
caps of one bulb and gently twist the bulb ¼ turn to remove. 
Using the same bulb, realign the two end pins with the slots 
on the bulb mounts. Push the pins back into the spring-load 
mounts and twist the bulb ¼ turn. Repeat process for each of 
the remaining bulbs.

Plug the ZEISS UVClean back into the electrical outlet and turn 
back on. Close the lid and press the Start button on top of the 
lid to begin a disinfection cycle. 

If the LED indicator still flashes red, proceed to Step 3.

Step 3 – Replace UVC Bulbs

Replace all UVC bulbs according to the instructions included 
in the ZEISS UVClean Ultraviolet-C (UVC) Eyeglass Disinfection 
Maintenance Kit.   

To order a ZEISS UVClean Ultraviolet-C (UVC) Eyeglass 
Disinfection Maintenance Kit, please contact your 
ZEISS Business Development Representative or call 
Customer Service at 1-800-201-4143.

CAUTION:

 Only use approved ZEISS UVC replacement 

bulbs. Never substitute or use other UV bulbs in the ZEISS 
UVClean. UVC disinfection dosage delivered by the device 
cannot be guaranteed if non-approved UV replacement 
bulbs are used. Additionally, usage of non-approved UV 
bulbs negates UL approval and compliance of the device.

Once all UVC bulbs are replaced, plug the ZEISS UVClean back 
into the electrical outlet. Turn the device back on. Close lid and 
press the Start button on top of the lid to begin a disinfection 
cycle. 

If the LED indicator returns to flashing red, proceed to Step 4.

Step 4 – Replace Device

A LED indicator light that continues to flash red indicates an 
electronic issue with the device. The ZEISS UVClean needs to 
be replaced.

If the device is under warranty, you will receive a replacement 
at no charge. Please contact your ZEISS Business Development 
Representative or call Customer Service at 1-800-201-4143 for 
replacement details.

Summary of Contents for UVClean

Page 1: ...ZEISS UVClean User Manual www zeiss com UVClean ...

Page 2: ...Table of Contents 3 General Information 4 Package Contents 5 About the ZEISS UVClean 6 Functionality 8 Safety and Operation 8 Cleaning and Recycling 9 Technical Information ...

Page 3: ... objects as severe damage to the device and or object and injury to the operator may result if misused What it is The ZEISS UVClean Uses UVC light for the safe and effective disinfection of eyeglass frames and lenses Is designed for the unique geometry of eyeglasses to prevent UVC shadows on frames lenses nose pads and temples Direct UVC illumination ensures thorough exposure which is critical for...

Page 4: ...4 Package Contents Device UVC Bulb kit and End caps Quick Start Guide Power cord ...

Page 5: ... the sides and in the middle of the bottom chamber Figure 3 Wearing the provided gloves align the two end pins with the slots on the bulb mounts Figure 4 Push the pins fully into the spring load mounts and twist the bulb turn The pins will click into place Repeat for the remaining three bulbs After installing the bulbs slide each cap outward toward the wall of the chamber Figure 5 Install all four...

Page 6: ...and while LED lights are pulsing blue Eye and skin damage may result if lid is opened while LED lights are pulsing blue LED indicator lights Lights along the underside of the lid illuminate colors Figure 8 and patterns to indicate operating conditions Solid Blue Ready to run a disinfection cycle Pulsing Blue Disinfection cycle running Solid Green Disinfection cycle complete Yellow Bulb replacement...

Page 7: ...ndicator still flashes red proceed to Step 3 Step 3 Replace UVC Bulbs Replace all UVC bulbs according to the instructions included in the ZEISS UVClean Ultraviolet C UVC Eyeglass Disinfection Maintenance Kit To order a ZEISS UVClean Ultraviolet C UVC Eyeglass Disinfection Maintenance Kit please contact your ZEISS Business Development Representative or call Customer Service at 1 800 201 4143 CAUTIO...

Page 8: ...fety thresholds Cleaning device Interior To prevent dust buildup within the chamber always keep the ZIESS UVClean lid closed when not in use Never spray or pour liquid cleaners inside the device Exterior The device exterior may periodically be wiped with a soft cloth and mild cleaner Never use harsh chemicals or spray cleaner directly onto the ZEISS UVClean Always spray cleaner onto soft cloth the...

Page 9: ...isionCare_USA Technical Information Specifications Bulbs 4 non ozone generating UVC lamps Dimensions 12 Wide x 12 High x 20 Deep Weight 20 2 lbs Power cord NEMA 5 15 to IEC 320 C13 for use with 120 volt electrical outlets UVClean electrical power input 120 VAC 50 60 Hz 330 W Listed by NRTL MET Canada USA 250V 10A CAUTION HIGH INTENSITY UV LIGHT DO NOT OPERATE WITH LID OPEN ...

Reviews: