Note: Acquiring more pixels than ”optimal” almost never makes sense, but feel free to decrease the 
pixel number to increase acquisition speed at the cost of resolution. 

10.

 

To capture single images, click at ‘Snap’.  

 

 
 

11.

 

Setting up Z-stack:   
 

-

 

Enable Z-Stack and the Z-Stack dialog will appear 

-

 

Press “Live” and find first and last slices with the focus knob and select them in the software accordingly. 

-

 

The software will suggest the optimal slice number and interval which you can change manually 

-

 

Press “Start experiment” 

 

12.

 

Setting up time series:  
 

-

 

Enable “time series” 

-

 

Time series dialog will open 

-

 

Set cycle number and interval time 

-

 

Press “start experiment” 

 

13.

 

Tile Scan (in this mode you have: Tiles, Multipositions) 
 

-

 

Enable “tile scan” 

-

 

Tile scan dialog will open 

-

 

Set desired number of horizontal and vertical tiles 

-

 

You can choose different regions to scan. Choose regions and press button “+” to add it. 

-

 

Press “start experiment” 
Advanced Set up allows you to do preview scan of region of interest. 
 

14.

 

Experiment designer. 
 

-

 

Enable “Experiment designer”.  

-

 

Experiment designer dialog will open 

-

 

Design your experiment using all options (z-stack, tiles  etc). Choose how many blocks you want to have.  You 
can add loops. 

-

 

Press “Start experiment” 

 

15.

 

Data Saving: 
 

Save your data to D: DATA folder in your own  folder. Note that this computer is not backed Up. Please move 
your data as soon as possible using portable storage devices or connect with intranet. 
You can also use “Auto save”, your experiment and snaps will go directly to your folder. 

 
 

V.

 

Switching off: 
 

1.

 

DO NOT save any “Experiment” or overwrite the defult “Experiment” (marked with a home logo) as this setting 
will be used in the next start up of the system and might cause potential damage to GaAsP PMTs. 

2.

 

Close the ZEN  Software 

3.

 

Shut down the computer. 

4.

 

Turn off the cool LED (5) 

Summary of Contents for LSM800-2

Page 1: ...nhole and pixel sizes as well as the sectioning thicknesses are all incorrect IMPORTANT LSM800 upright has two new GaAsP PMT detectors As already explained during intros the new PMT are very sensitive and easy to oversaturate and damage GaAsP detectors are shutting down if oversaturated terminating your experiment without further warnings Please use them correctly ...

Page 2: ...ive Gallium arsenide phosphide photomultiplier tube GaAsP PMTs detectors with free choice of the spectral range Uses a tunable short pass and bandpass filter in front of the PMTs Be very careful with using it to prevent instrument damage If oversaturated detectors will switch off automatically An ESID detector for transmitted light Simultaneous detection of only two dyes possible more than two hav...

Page 3: ...equential best signal scan press Apply after choosing the appropriate setup 4 If you are using the detector for transmitted light activate the ESID Detector represented in the light path scheme Make sure Köhler illumination is set properly if you are getting bad bright field images 5 Go to Live acquisition modus and find the focus manually 6 Set pinhole to Airy 1 in the channel dialog to achieve c...

Page 4: ...es You can choose different regions to scan Choose regions and press button to add it Press start experiment Advanced Set up allows you to do preview scan of region of interest 14 Experiment designer Enable Experiment designer Experiment designer dialog will open Design your experiment using all options z stack tiles etc Choose how many blocks you want to have You can add loops Press Start experim...

Page 5: ...5 Switch off COMPONENTS 3 6 Switch off SYSTEM 2 7 Switch off multiple sockets 1 ...

Page 6: ...ser should be avoided at all times Laser safety goggles are situated at all workspaces and should be used in any situation where potential contact of eyes with Laser light of the classes 3B or 4 is possible according to the previous two points Laser safety goggles have to be worn at all times of operation of Laser Class 4 Laser safety goggles have to be worn at all times of operation laser Class 3...

Reviews: