background image

7.

 

Login: LSMuser (password: Useme!11) 

8.

 

Open  

ZEN Blue  

Software and press “ZEN System” 

9.

 

“Image Processing” is for offline analysis of your images, please do not use it here 
 

 
 

III.

 

Visual examination of the sample 
 
1.

 

Put mechanical switch on the right hand side of the microscope to the ”Eye” position. 

2.

 

Press “Locate” in the main microscope menu. 

3.

 

Since this is an upright system, changing objectives can only be manually done by rotating the objective 
revolver. Rotation should work very easily when objectives are in the “up” position. 

4.

 

For imaging or visual inspection, the objectives must be lowered entirely with the black wheel in front of 
the microscope stand. 

5.

 

Please make sure that the chosen objective corresponds to the one shown in the software. If this is 
not the case, please restart the system. (Otherwise, your pinhole and pixel sizes, as well as the 
sectioning thickness are all incorrect.) 

6.

 

If you wish to view your sample in Brightfield mode, press the “TL” button 

7.

 

For visual examination of fluorescence, press the buttons DAPI, GFP or RFP 

 

IV.

 

Acquiring confocal images 
 
1.

 

Set mechanical switch on the right hand side of the microscope to the “Camera” position. 

2.

 

Select “Acquisition” in the main menu of Zen. (Software will ask you about the mechanical switch, 
Press”OK”.)   

3.

 

Go to “Smart Setup” add your dyes pressing “+” button (set the LUT) and configuration, you can choose 
between simultaneous (fastest) and sequential (best signal) scan, press ”Apply” after choosing the 
appropriate setup. 

4.

 

If you are using the detector for transmitted light: activate the ESID Detector represented in the light 
path scheme. Make sure Köhler illumination is set properly if you are getting bad bright field images! 

5.

 

Go to “Live” acquisition modus and find the focus manually. 

6.

 

Set pinhole to Airy 1 in the channel dialog to achieve confocality. 

7.

 

Press the “Auto exposure” button and the software will adjust the gain of the PMTs but not the laser 
power, according to the brightness of your dyes. 
7.1.

 

 What Auto exposure does is to increase the PMT voltages until about 1% of all pixels are 
overexposed. Gain cannot go below 500 V and above 900V. 

8.

 

In the channel dialog you can adjust the pinhole, gain and laser power manually to have the best image. 
You can increase laser power by default only till 4,5%. If you need more power you need to enable 
“Higher Intensity Laser  Range”. 

Be aware that GaAsP PMT are very sensitive and if you will put to much 

power, they shut down for the course of imaging (& do not turn on automatically). In that case you need 
to stop acquisition and restart the imaging manually!

 

9.

 

To zoom in your image, go to “Acquisition parameteres” at the drop down menu. With “Crop” you can 
select the desired area of your image, then go to “Live” or “Snap” to apply, “Reset all” will remove the 
zoom again. Depending on the scan speed, the lowest zoom is 0.5 and not 1.0. Before taking an image, 
press “optimal” in the “acquisition mode” to adjust the optimal number of pixels and the pixel size 
according to Nyquist sampling criteria. 

Summary of Contents for LSM800-2

Page 1: ...nhole and pixel sizes as well as the sectioning thicknesses are all incorrect IMPORTANT LSM800 upright has two new GaAsP PMT detectors As already explained during intros the new PMT are very sensitive and easy to oversaturate and damage GaAsP detectors are shutting down if oversaturated terminating your experiment without further warnings Please use them correctly ...

Page 2: ...ive Gallium arsenide phosphide photomultiplier tube GaAsP PMTs detectors with free choice of the spectral range Uses a tunable short pass and bandpass filter in front of the PMTs Be very careful with using it to prevent instrument damage If oversaturated detectors will switch off automatically An ESID detector for transmitted light Simultaneous detection of only two dyes possible more than two hav...

Page 3: ...equential best signal scan press Apply after choosing the appropriate setup 4 If you are using the detector for transmitted light activate the ESID Detector represented in the light path scheme Make sure Köhler illumination is set properly if you are getting bad bright field images 5 Go to Live acquisition modus and find the focus manually 6 Set pinhole to Airy 1 in the channel dialog to achieve c...

Page 4: ...es You can choose different regions to scan Choose regions and press button to add it Press start experiment Advanced Set up allows you to do preview scan of region of interest 14 Experiment designer Enable Experiment designer Experiment designer dialog will open Design your experiment using all options z stack tiles etc Choose how many blocks you want to have You can add loops Press Start experim...

Page 5: ...5 Switch off COMPONENTS 3 6 Switch off SYSTEM 2 7 Switch off multiple sockets 1 ...

Page 6: ...ser should be avoided at all times Laser safety goggles are situated at all workspaces and should be used in any situation where potential contact of eyes with Laser light of the classes 3B or 4 is possible according to the previous two points Laser safety goggles have to be worn at all times of operation of Laser Class 4 Laser safety goggles have to be worn at all times of operation laser Class 3...

Reviews: