08/2007
21
Switching off the system
"
Click on the
File
button in the
Main
Menu
bar and then click on the
Exit
button to leave the
ZEN
software.
"
If any lasers are still running you should shut them off now in the pop-up window indicating the lasers
still in use.
"
Shut down the computer.
"
Wait until the fan of the Argon laser has switched off.
"
On the REMOTE CONTROL turn off the
Components
switch and the
System/PC
switch (Fig. 1).
"
Switch off the X-Cite 120 lamp or the HBO 100 mercury burner.
"
Switch off the UV-Ar laser of by the toggle switch on the power supply (Fig. 2).