5 First Operating Steps | 5.6 Finding Appropriate Detector Settings
Detector
Pressure mode
The variable pressure mode can either be normal VP or extended EP. This depends on the fac-
tory configuration. If EP is enabled the VP is replaced by EP.
For information on special set-up procedures for the detectors, refer to:
§
§
§
Setting up the C2DX Detector [
§
§
Setting up the YAG BSD Detector [
Procedure
1. Select the Detectors tab of the SEM Controls panel.
2. Select the detector from the
Signal A
drop-
down list.
5.6.2 Setting up the SE Detector
Procedure
1. In the
Detectors
tab of the SEM Controls panel, select
SE2
from the
Signal A
drop-down
list.
2. Use the
Collector Bias
scroll bar to adjust the
collector bias.
INFO:
The default value is 300 V.
5.6.3 Setting up the C2D Detector
image of radiolaria
Instruction Manual ZEISS EVO | en-US | Rev. 10 | 354706-0780-006
83