5 First Operating Steps | 5.6 Finding Appropriate Detector Settings
Info
The C2DX detector requires VP pressures of at least 20 Pa to acquire a good signal.
A good starting point is EHT = 20 kV, I Probe = 500 pA, Chamber pressure = 30 Pa,
WD = 8.5 mm, and scan speed = 5.
Procedure
1. In the SEM Controls panel, select the
Detectors
tab.
2. From the
Signal A
drop-down list, select
C2DX
.
3. In the
Panel Configuration Bar
, select
C2DX Control
.
à
The
C2DX Control
panel is displayed.
4. Select
C2DX Gain = Low
.
5. Adjust the
C2DX Bias
scroll bar until you obtain a good and stable image.
INFO:
A typical value for
C2DX Bias
is 50 %. Contrast and brightness should also be ad-
justed to obtain an image.
6. Click
C2DX Auto Level
.
7. If you are unable to reach saturation or if the image quality is not good, select
C2DX Gain
= High
and repeat the procedure.
5.6.5 Setting up the BSD Detector
Prerequisite
ü
Chamber door is open.
Procedure
1. Check that the lens-mounted BSD detector is fitted.
2. If it is in the parked position, remove the BSD
detector from the parked position.
Instruction Manual ZEISS EVO | en-US | Rev. 10 | 354706-0780-006
85