2.7
Attachment of condenser
•
Remove condenser, e.g. swing-in version 0.9 Z Pol, from the packaging.
•
Carefully move microscope stage and stage carrier D to the uppermost stop position via the coarse
drive (2-13/
1
). Caution: risk of collision with objectives !
•
Lower the condenser holder via the condenser drive (2-13/
2
) and loosen the clamping screw (2-13/
4
).
Fold down condenser front lens via the lever (2-13/
5
) and push condenser into the guiding fork from
the front.
☞
The notch (2-13/
3
) in the condenser carrier is used for condenser orientation. The swing-in
condenser 0.9 Z Pol must be oriented in such a way that the lever (2-13/
5
) points horizontally
to the left.
Fig. 2-13
Attachment of condenser