Carl Zeiss
Axiotech and Axiotech
vario
2-10
B 40-020 e 06/99
2.5.2
Compensation of ametropia when eyepiece reticles are used
The correct use of an eyepiece reticle requires two focusing eyepieces, e.g. W-PL 10
×
/23 Br. foc., to
enable compensation of ametropia.
•
Use the eyelens of the focusing eyepiece to focus on the line figure of the eyepiece reticle; focus on
the edge of the field of view if no eyepiece reticle is used.
•
Focus on the microscope image of a specimen via the coaxial focusing drive by looking through the
eyepiece with reticle.
•
When the image and the eyepiece reticle are in focus in the above eyepiece, focus the image for the
second eye via the focusing eyelens of the second eyepiece.
☞
The position of the focusing drive on the stand must not be moved.
2.6
Setting of eyepiece (interpupillary) distance and viewing height
•
The eyepiece distance is matched to the user's
individual interpupillary distance by swinging
the two eyepiece tubes symmetrically towards
one another (2-12).
Fig. 2-12
Setting of eyepiece distance and
viewing height of the binocular
tube