zebris Medical GmbH
Canid Gait Technical Specifications and User Manual
Page 15/33
4.6.2 Synchronization Output (SYNC-OUT)
If a third party device is connected to the synchronization output SYNC-OUT then depending on
the setting of the configuration window of the application software, a CANID GAIT system con-
trolled, will trigger a synchronized measurement of the third party device either via start/ stop or
“sample by sample” mode.
Electrical Specifications
Output Resistance
100
V
OH
(High-Level Output Voltage)
≥ 2.0 V
V
OL
(Low-Level Output Voltage)
≤ 0.8 V
Integrated LEMO socket in the interspace box
Series
"00"
, four pole, Coding 0°
LEMO-Part-No.: EPG.00.304.NLN
View
Socket, Front Side
Socket Coding: 0°
Suitable Plug for SYNC-OUT
LEMO-Part-No.: FGG.00 304.CLADxxxx
Pin-Assignment
Pin 1
+5 V
Pin 2
GND
Pin 3
Activ_OUT
Pin 4
Clk_OUT
View
Plug, Solder Side
Plug Coding: 0°