•  The temperature of the airflow for breathing produced by CPAP devices can be higher than 

the temperature of the room. This is true of all CPAP devices. Caution should be exercised 
when using your 

Z1

 at room temperatures warmer than 90˚F (32˚C).

•  Only operate the 

Z1

 and accessories within the operating temperature, humidity and 

altitude conditions specified in the Technical Specifications section of this User Guide.

•  Only store or transport the Z1 and accessories within the storage and transport 

temperature, humidity and altitude conditions specified in the Technical Specifications 
section of this User Guide.

•  Damage to the 

CPAP Tube

 may result in air leaks. If this occurs, stop using the damaged 

CPAP Tube

 and replace immediately.

Battery Warnings and Cautions

The 

PowerShell

 comes with a removable 

Lithium Ion Battery

.  Careful attention must be paid to 

the warnings and cautions described below relating to the 

Battery

.  Misusing the 

Battery

 may 

cause the 

Battery

 to get hot, rupture, ignite and cause serious injury.

 

WARNINGS: 

•  Do not place the 

Battery

 in fire or heat the 

Battery

.

•  Do not install the 

Battery

 backwards so that the polarity is reversed.

•  Do not connect the positive terminal and the negative terminal of the 

Battery

 to each other 

with any metal object (such as wire).

•  Do not carry or store the 

Battery

 together with necklaces, hairpins, or other metal objects.

•  Do not pierce the 

Battery

 with nails, strike the 

Battery

 with a hammer, step on the 

Battery

 or 

otherwise subject it to strong impacts or shocks.

•  Do not solder onto the 

Battery

.

•  Do not expose the 

Battery

 to water or salt water, or allow the 

Battery

 to get wet.

•  Do not disassemble or modify the 

Battery

. The 

Battery

 contains safety and protection 

devices which, if damaged, may cause the 

Battery

 to generate heat, rupture or ignite.

•  Do not place the 

Battery

 on or near fires, stoves or other high-temperature locations. Do not 

place the 

Battery

 in direct sunlight, or use or store the 

Battery

 inside cars in hot weather. 

Doing so may cause the 

Battery

 to generate heat, rupture or ignite. Using the 

Battery

 in this 

manner may also result in a loss of performance and a shortened life expectancy.

•  Immediately discontinue use of the 

Battery

 if, while using, charging or storing the 

Battery

the 

Battery

 emits an unusual smell, feels hot, changes color, changes shape or appears 

abnormal in any other way. Contact the Human Design Medical Service Center at 1-855-
436-8729 [1-855-HDMUSA9] if any of these problems are observed.

•  Do not place the 

Battery

 in microwave ovens, high-pressure containers, or on induction 

cookware.

Summary of Contents for CPAP System

Page 1: ...Human Design Medical LLC Z1 CPAP System CONTINUOUS POSITIVE AIRWAY PRESSURE SYSTEM User Guide User Guide English...

Page 2: ......

Page 3: ...Box 10 Assembling and Powering Up 11 Using Your Z1 13 LCD Display 13 Starting and Stopping Treatment 14 Programming Ramp Mode 15 Inserting Removing and Replacing the Micro SD Data Card 15 User Softwa...

Page 4: ...ng portable CPAP device Please read this User Guide carefully before using your Z1 Owner User Responsibility The owner user of the Z1 CPAP System is responsible for reading and understanding this User...

Page 5: ...conditions Severe bullous lung disease Pneumothorax Pathologically low blood pressure Dehydration Cerebrospinal fluid leak recent cranial surgery or trauma Adverse E ects Users should report unusual c...

Page 6: ...masks and connectors recommended by your physician clinician respiratory therapist or health care provider Your Z1 is intended to be used with CPAP masks or connectors which have vent holes to allow c...

Page 7: ...P Tube and or air inlet of your Z1 while in operation could lead to overheating of the Z1 If you put the Z1 on the floor make sure the area is free from dust and clear of bedding clothes or other obje...

Page 8: ...m or at least 8 feet 2 5 m above the patient It must also comply with the relevant test standard For personal computers the international standard is IEC 60950 or equivalent CAUTIONS Be careful not to...

Page 9: ...inal and the negative terminal of the Battery to each other with any metal object such as wire Do not carry or store the Battery together with necklaces hairpins or other metal objects Do not pierce t...

Page 10: ...se the Battery to power any device other than the Z1 Only use or charge the Battery within the operating temperature humidity and altitude conditions specified in the Technical Specifications section...

Page 11: ...ses the Filter For proper operation of the Z1 the Filter should be replaced monthly or more often if necessary Filter The Z1 uses a specially designed filter The material has been specifically chosen...

Page 12: ...packaging 3 Carefully inspect all contents against the Package Contents list on page 8 to assure all components are present Keep in mind that your health care provider might have completed some of the...

Page 13: ...hrough the Filter and no air bypasses the Filter 4 Press the End Cap onto the Z1 Figure 1 NOTE To remove the End Cap squeeze firmly on the sides of the End Cap and pull it away from the Z1 Tube Adapte...

Page 14: ...er the device is receiving power PowerShell and Battery Available Separately or as Part of Z1 Unplugged WARNING The Z1 must only be used with the optional Human Design Medical PowerShell with Human De...

Page 15: ...within 2 3 days Using Your Z1 LCD Display Treatment Pressure Display The current treatment pressure is shown on the LCD display Once set the prescribed treatment pressure will be retained in the Z1 s...

Page 16: ...R is displayed if there is excessive air leakage See Reminder Alert page 18 for an explanations of REMINDER conditions Starting and Stopping Treatment 1 Press the Start Stop Button on the Z1 to initia...

Page 17: ...Replacing the Micro SD Data Card Optional Programming Ramp Mode Some users find it comforting to acclimate to CPAP therapy each session This can be accomplished by programming the Ramp Mode which all...

Page 18: ...e Charging the Battery PowerShell Only WARNING Only charge the Battery with the supplied Human Design Medical PowerShell and the supplied Human Design Medical Power Supply Do not use any other battery...

Page 19: ...ure they are dry before plugging back in CAUTION Do not use bleach chlorine alcohol or aromatic based solutions moisturizing or antibacterial soaps or scented oils to clean the Z1 Never attempt to cle...

Page 20: ...e Z1 The Z1 is designed to only work with approved materials A filter which is not approved may cause damage to the Z1 from overheating CAUTION Do not operate the Z1 without the Filter The Z1 must ope...

Page 21: ...upply is connected and the power outlet has power Loose connection of the Power Supply at the Z1 or loose connection of the Power Cord at the AC Adapter Check the connections Z1 is not locked into pos...

Page 22: ...re not positioned correctly and mask is leaking Adjust position of the mask and headgear until leak ceases Incorrect CPAP Tube selected Do not use tubing which was not recommended by your clinician Ch...

Page 23: ...andby CPAP and CPAP with ramp Operating pressure range 4 to 20 cm H2O Maximum single fault steady pressure state 30 cm H2O Deliverable pressure accuracies 0 6 cm H2O 4 of the measured reading Sound pr...

Page 24: ...out by any service organization that has not been expressly authorized by Human Design Medical to perform such repairs c any damage or contamination due to cigarette pipe cigar or other smoke d damag...

Page 25: ...provided they have been tested for Electromagnetic Compatibility EMC in accordance with the current version of RTCA DO 160 Section 21 Category M Human Design Medical has successfully completed testing...

Page 26: ......

Page 27: ...St Suite P Charlottesville VA 22902 About you NAME First MI Last EMAIL ADDRESS MAILING ADDRESS Street or PO Box City State Zip PHONE NUMBER About the device WHAT DEVICE DID YOU PURCHASE WHERE DID YOU...

Page 28: ......

Page 29: ......

Page 30: ...Human Design Medical LLC 119 Braintree Street Suite 703 Boston MA 02134 Z1 User Guide v1 11012013 40 0007 Rev A...

Reviews: