• The temperature of the airflow for breathing produced by CPAP devices can be higher than
the temperature of the room. This is true of all CPAP devices. Caution should be exercised
when using your
Z1
at room temperatures warmer than 90˚F (32˚C).
• Only operate the
Z1
and accessories within the operating temperature, humidity and
altitude conditions specified in the Technical Specifications section of this User Guide.
• Only store or transport the Z1 and accessories within the storage and transport
temperature, humidity and altitude conditions specified in the Technical Specifications
section of this User Guide.
• Damage to the
CPAP Tube
may result in air leaks. If this occurs, stop using the damaged
CPAP Tube
and replace immediately.
Battery Warnings and Cautions
The
PowerShell
comes with a removable
Lithium Ion Battery
. Careful attention must be paid to
the warnings and cautions described below relating to the
Battery
. Misusing the
Battery
may
cause the
Battery
to get hot, rupture, ignite and cause serious injury.
WARNINGS:
• Do not place the
Battery
in fire or heat the
Battery
.
• Do not install the
Battery
backwards so that the polarity is reversed.
• Do not connect the positive terminal and the negative terminal of the
Battery
to each other
with any metal object (such as wire).
• Do not carry or store the
Battery
together with necklaces, hairpins, or other metal objects.
• Do not pierce the
Battery
with nails, strike the
Battery
with a hammer, step on the
Battery
or
otherwise subject it to strong impacts or shocks.
• Do not solder onto the
Battery
.
• Do not expose the
Battery
to water or salt water, or allow the
Battery
to get wet.
• Do not disassemble or modify the
Battery
. The
Battery
contains safety and protection
devices which, if damaged, may cause the
Battery
to generate heat, rupture or ignite.
• Do not place the
Battery
on or near fires, stoves or other high-temperature locations. Do not
place the
Battery
in direct sunlight, or use or store the
Battery
inside cars in hot weather.
Doing so may cause the
Battery
to generate heat, rupture or ignite. Using the
Battery
in this
manner may also result in a loss of performance and a shortened life expectancy.
• Immediately discontinue use of the
Battery
if, while using, charging or storing the
Battery
,
the
Battery
emits an unusual smell, feels hot, changes color, changes shape or appears
abnormal in any other way. Contact the Human Design Medical Service Center at 1-855-
436-8729 [1-855-HDMUSA9] if any of these problems are observed.
• Do not place the
Battery
in microwave ovens, high-pressure containers, or on induction
cookware.