background image

DP-2010L Version 04032001 ATEX Rev.

YZ Systems, Inc. • 3101 Pollok Drive • Conroe, Texas • USA • 77303 • P: 936.788.5793 • F: 936.788.5720

31

Diagram #9:  DuraSite Portable Sample Vessel Instructions

Purpose:  The DuraSite Portable Sample Vessel permits the
user to remove a liquid or gas hydrocarbon sample from a
pipeline or a sampling device.  This is accomplished without
changing the pressure of the product or exposing it to a
contaminant fluid.  If properly used and maintained the Dur-
aSite will provide many years of safe, accurate and clean
sampling.

Use:  The DuraSite is a very safe device to use.  As with any
equipment dealing with flammable products, it is mandatory
that a good, thorough operator training procedure be estab-
lished prior to use.

Typical use of the cylinder would be as follows:

Step 1:  (In The Lab)  Connect a regulated inert gas supply to
the pre-charge valve.  The product valve should be open.  By
carefully controlling the pre-charge valve and the regulator, the
cylinder can be slowly charged with pre-charge gas (NOTE:
This should be done slowly to prevent slamming the piston
down to the opposite end).  The pressure on the pre-charge
pressure gauge should be brought to a reading of 10-50 psi
above the expected pressure of the product in the field .  Close
the pre-charge valve and disconnect the gas supply.  Check the
pre-charge valve, relief device, and the pre-charge pressure
gauge for leaks.  Any leaks should be stopped before continu-
ing.  The vessel should be placed in a padded carrying case and
made ready for field use.
Proceed to 

EITHER Step 2, or Step 3 as

required for your application.

STEP 2: FOR COLLECTION OF SAMPLE VIA SPOT SAMPLE  OR
FROM  COMPOSITE  ACCUMULATOR  VESSEL.

2a:  Connect the product end of the pre-charged sample vessel
to the product supply.
(Sampler product removal valve, or Pipeline sample probe)
NOTE:  the pre-charge pressure gauge reading should be
greater than the product supply pressure reading.  If not,
repeat Step 1 above.

2b:  Once the vessel is connected to the product supply, it is
necessary to vent a small amount of product prior to filling the
vessel.  This assures fresh product and removes any air or gas
when dealing with liquids.  This can be done by loosening the
product purge valve a very small amount until the product is
purged.  After thorough purging, the product purge valve should
be tightened.

2c: The product pressure gauge reading should be 10-50 psi
below the pre-charge pressure gauge reading. By carefully
opening the pre-charge valve, the pressure becomes equal-
ized, then begins to drop below the product pressure.  The pre-
charge valve should be carefully controlled so as to not vent the
pre-charge gas too fast.

2d:  When the cylinder becomes a maximum of 80% full (see
volume indicator), all valves should be closed.  The product
connection is slowly broken in order to vent any trapped product.
After vessel removal, all connections should be checked for
leaks and the pre-charge and product valve ports capped to
prevent leakage.

2e:  Pack the DuraSite in appropriate carrying case to meet
D.O.T. guideline, with D.O.T.
paperwork and transport to lab for analysis.

STEP 3: FOR DIRECT CONNECTION TO
SAMPLER.

3a:  Connect the sampler discharge port to the product inlet port
to the DuraSite using 1/8" stainless steel tubing.

3b:   (Gas sampling)  Connect the pre-charge port to the
DuraSite to the pipeline for pre-charge pressure (Proceed to
step 3d), or configured like the liquid sample application below.
(Step 3c)

3c:  (Light sampling)  Pre-charge the DuraSite as indicated in
Step 1, then install a pressure relief valve to the pre-charge port
and open the pre-charge valve on the DuraSite. (The pressure
relief valve should have a relief pressure setting of  approxi-
mately 100 psi above line pressure.)

3d:  Open the product inlet valve of the DuraSite and the purge
valve on the sampler. Next open the purge valve on the product
end of the DuraSite and allow product to purge all lines and
connections out.

3e:  Close purge valves and begin sample cycle.

3f:  At the end of sample cycle, close product inlet valve on the
DuraSite and remove the DuraSite. Pack the DuraSite in
appropriate carrying case to meet D.O.T. guideline, with D.O.T.
paperwork and transport to lab for analysis.

Step 4:  (In The Lab)  Prior to analysis, the product should be
mixed.  This is accomplished simply and efficiently by inverting
the cylinder end-over-end,  causing the mixing ball to fall
through the product.  Approximately 10-12 trips of the mixing ball
through the product assures a homogenous solution.

Step 5:  The regulated pre-charge gas should be reconnected
to the pre-charge side of the cylinder.  The pre-charge gas
supply should remain open during analysis.

Step 6:  Purging a small amount of product from the vessel
removes unmixed product from the tee, relief device, gauge,
etc.  The unit can now be connected to a chromatograph and
the product analyzed.

Step 7:  After analyzing, the remainder of the product should be
dumped and the vessel properly cleaned.  Normal cleaning can
be accomplished by rinsing the product end with a petroleum
solvent and flushing with acetone.  If a more thorough cleaning
is required, the vessel should be disassembled.

WARNING:  A portable sample vessel should never be  filled
to more than 80%.  This allows a 20% pre-charge cushion to
absorb thermal expansion of the product.
Shipping:  Extreme care should be taken when preparing a
vessel for shipment.  Both valves should be capped to prevent
possible leakage.  The vessel should be placed in a
snug-fitting, well-padded and durable case.  All applicable DOT
regulations should be adhered to.

Summary of Contents for DynaPak 2010L

Page 1: ...DynaPak2010L SystemSupport Manual Version04032001...

Page 2: ...d the customer s exclusive remedy will be replacement of the product s or at YZ s option YZ s reasonable effort to make the product meet the warranty set forth herein YZ disclaims all other warranties...

Page 3: ...tact your YZ representative or YZ Customer Service at 936 788 5593 This sampler can operate at pipeline pressures from 10 psig 69 Bar up to 1500 psig 103 4 Bar 1949 Factory System Approved Mutual Samp...

Page 4: ...nd resume operation C Proportional to flow sampling in this mode of operation the Z 65 counter operates as a dividing counter The Z 65 counter periodically energizes a low power solenoid valve As in t...

Page 5: ...could cause aerosols or significant pressure drops 4 2 The sampler should not be located within the defined meter tube region AGA 3 manual A The number of unobstructed straight pipe diameters upstrea...

Page 6: ...quid leak detector CAUTION Incorrect operation of valves over tightening can result in damage to the valve components isolation valve bonnet assembly which might result in the valve stem being screwed...

Page 7: ...or the high pressure supply valve without ensuring that the equalization valve is open Failure to do so may damage the DPS 2 s internal components d Open the other supply valve e Close the equalizatio...

Page 8: ...created 6 2 Variable volume constant pressure cylinder The free floating piston cylinder DuraSite may be installed in a horizontal position on an optional vessel rack Free floating piston cylinders sh...

Page 9: ...Bar 7 3 Turn the stroke adjustment knob on the top of the pump counterclockwise to set the pump displacement at 4cc stroke 7 4 Move all of the mode switches on the Z 65 to their off positions 7 5 Mov...

Page 10: ...E The time 18 minutes above corresponds to the dial setting shown for the Z65 model with the timer range setting in the factory position jumper on the two left pins See section 12 4 Timer Range Settin...

Page 11: ...8 6 Turn mode switch 3 to on 8 7 Turn mode switch 4 to off 8 8 Press the test button once to initiate the timer sequence 8 3 Adjust the pump volume adjustment knob to the value used in the calculation...

Page 12: ...time 18 minutes above corre sponds to the dial setting shown for the Z65 model with the timer range setting in the factory position jumper on the two left pins See section 12 4 Timer Range Setting 3 2...

Page 13: ...9 6 Turn mode switch 3 to off 9 7 Turn mode switch 4 to off 9 8 Press the test button once to initiate the timer sequence NOTE For the Z 65 to operate the DP 2010L System a minimum pressure different...

Page 14: ...ng connections to the Z 65 terminal strip see illustration b Turn mode switch 1 to on c Turn mode switch 2 to off d Turn mode switch 3 to on e Turn mode switch 4 to off OR 10 3 If the input is a volta...

Page 15: ...ber of pulses you want to count before the sample pump strokes Example 20 pulses turn dials to 20 Press the test button once to load the value into the memory NOTE If the calculated counter setting is...

Page 16: ...p body d Screw the stroke adjustment knob all the way down to the 0 cc stroke setting e Unscrew the pump body by hand from the inlet check valve assembly Separation at this point is recommended to mai...

Page 17: ...l o rings and bushings to prevent damage j Install the body bushing into the bottom of the pump body NOTE Apply a light coat of assembly grease on the plunger shaft prior to installation k Insert all...

Page 18: ...te if the battery is low NOTE The solenoid must be connected to test the battery condition Battery condition cannot be tested with a volt meter Replacing a Depleted Battery 1 Remove the four thumb scr...

Page 19: ...for the DynaPak 2000 Series gas samplers Part Number Description Qty Location C4 0004 Filter element 1 see diagrams 3 and 4 D3 0002 DP 2000 pump seal kit 1 see diagrams 1 and 2 D3 0003 YZ filter regul...

Page 20: ...ke counter set the mode switches as follows 1 Positions 1 2 and 3 on 2 Position 4 off B Set the time switches to 00 C Unscrew the thumbscrews and remove the six position terminal strip and cover This...

Page 21: ...osition factory setting b x x minutes set the configuration jumper to the center position 2 Z 65 6 03 Range Setting a x x minutes set the configuration jumper to the far left position factory setting...

Page 22: ...te again This will normally occur very quickly and give the appearance that the green LED blinks on when the pulse input is received and removed 2 Voltage Pulse Input move mode switch 4 to the on posi...

Page 23: ...DP 2010L Version 04032001 ATEX Rev YZ Systems Inc 3101 Pollok Drive Conroe Texas USA 77303 P 936 788 5793 F 936 788 5720 23 Diagram 1 DP 2000 pump assembled...

Page 24: ...ustment Detent per Diaphragm Housing Diaphragm P N B1 0003 P N C3 0005 P N B1 0030 Set Screw P N C0 0096 P N C0 0014 Cap Screw 6 Each P N B1 0020 P N C0 0026 Inlet Screen P N C4 0006 P N C3 0008 Inlet...

Page 25: ...DP 2010L Version 04032001 ATEX Rev YZ Systems Inc 3101 Pollok Drive Conroe Texas USA 77303 P 936 788 5793 F 936 788 5720 25 Diagram 3 YZ filter regulator assembled...

Page 26: ...2 Upper Dart P N A3 0071 Upper Spring P N C3 0010 P N A3 0070 Piston P N A3 0069 Piston Spring Spring Cap P N A3 0068 P N A3 0067 Bonnet Pneumatic Fitting P N A1 0113 P N A8 0016 Adjustment Screw P N...

Page 27: ...17 P N C0 0099 P N Varies See Above Pressure Gauge Stainless Steel Set Screw Inlet Rupture Disc 1800 psi P N A3 0083 Rupture Disc Retainer P N A3 0084 YZ Valve Body P N A3 0137 0 1000 PSI LinkPlus Mod...

Page 28: ...048 Qty 8 P N A6 0013 P N G2 0002 Cable Assembly 30 Ft Shrink Wrap SS Bolt P N C0 0047 Qty 8 P N H2 0002 Qty 2 Magnet Support Body Diaphragm Body Diaphragm Body P N A9 5003 High Pressure P N A9 5004 P...

Page 29: ...3 0142 1 12 2 Fuses Per Kit Ref No Description Part No Qty Z 65 6 Controller Assembly 1 1 Model Z 65 6 1 F2 0001 Model Z 65 6 03 F2 0018 2 Battery Pack E3 2001 1 3 Stroke Counter Assembly G1 0001 1 4...

Page 30: ...30 Diagram 8 Z 65 Installation Notes Wiring Control Documentation System Installation Notes and Recommendations...

Page 31: ...der becomes a maximum of 80 full see volume indicator all valves should be closed The product connection is slowly broken in order to vent any trapped product After vessel removal all connections shou...

Page 32: ...0 P N C6 1603 DS 1000 Gauge P N A3 0083 Rupture Disc P N A3 0084 YZ UniValve P N C6 1001 Pre Charge Head P N A3 0095 Valve Stem Assembly P N A3 0062 Seat Retaining Washer P N A3 0063 Valve Seat P N C6...

Page 33: ...DP 2010L Version 04032001 ATEX Rev YZ Systems Inc 3101 Pollok Drive Conroe Texas USA 77303 P 936 788 5793 F 936 788 5720 33 Notes...

Reviews: