Environment 
conditions

Temperature 

O5

~ 35

(

+41

~ 95

℉ )

Non-

condensing
Transport and storage:-20

~ 70

(-4

158

℉ )

Non-condensing

Humidity 

Operate: relative humidity 15%-90% Non-
condensing
Transport and storage: relative humidity 15%-
90% Non-condensing

Atmospheric pressure range

700 hPa~1060 hPa

Altitude

≤3000 m

Protection class

Type of Protection Against Electric Shock: Class 

 

Degree of Protection Against Ingress of Water: IP21 (the device)
                                                                           IP22 (the Power adapter)
Degree of Protection Against Electric Shock: Type BF Applied Part

Mode of 
operation

Continuous operation 

Maximum single 
fault steady 
pressure

Device will shut down in the presence of a single fault if the steady state 
pressure exceeds: 40 cmH

2

O

Sound 

Sound pressure level

Sound pressure level measured according 
to ISO 80601-2-70:2015 (CPAP mode) : 
28±2  dB (A)

Sound power level

Sound power level measured according to 
ISO 80601-2-70:2015 (CPAP mode) : 36±2 
dB (A)

Physical 
properties

Dimensions 
(length*width*height) 

285 mm*155 mm*125 mm or 
11.2'' * 6.1'' * 4.9''

Weight 

About 1500 g

Heated tubing

Plastic hose, about 1.8 m

Maximum volume of 
water tank

260 ± 10 mL 

Air outlet

22 mm (complies with ISO 5356-1:2015)

Temperature

Maximum temperature 
of heater plate 

55℃ ±4℃ ( 131℉ ±14.4℉ )

Cut-out

110

℃(

if it is damaged, return to manufacturer)

Maximum gas temperature

43

Air filter 

Material: Polyester non-woven fiber 
Average arrestance: ≥ 85% for~2.5 micron dust

Therapy pressure  4~20 hPa (adjustable, step is 0.5 hPa) or 4~20 cmH

2

O

Initial pressure

4~20 hPa (adjustable, step is 0.5 hPa) or 4~20 cmH

2

O

Maximum 
pressure 

4~20 hPa (adjustable, step is 0.5 hPa) or 4~20 cmH

2

O

Minimum 
pressure 

4~20 hPa (≤ maximum pressure, adjustable, step is 0.5 hPa) or 4~20 cmH

2

O

Ramp

0~45 minutes (adjustable, step is 5 minutes) 

Gas leakage 

2 L/min at the maximum rated pressure

Maximum 

flowrate

The performance of the device at set pressure according to ISO 80601-2-
70:2015 is shown below

Test pressures(hPa)

4

12

16

20

Average flow at the PATIENT-

CONNECTION PORT (L/min)

≥100

≥100 ≥100

≥100

≥100

Pneumatic flow path:

Atmosphere

Filter

Blower

Air tube

Flow sensor

Pressure sensor

Leak

Mask

Humidifier 

performance 

Humidification system output: ≥ 12 mg/L 
(The humidifier belongs to Category 2 according to the standard ISO 80601-2-

74:2017)

Relative humidity: ≥ 50%

Summary of Contents for YH-480A

Page 1: ...ay Pressure Units Suzhou Yuyue Medical Technology Co Ltd No 9 Jinfeng Road Suzhou Science Technology Town 215163 Suzhou Jiangsu PRC TEL 86 0512 67373001 web address www yuyue com cn IFU V 68 version 0...

Page 2: ...uction 03 3 Scope of application 03 4 Contraindications and adverse effects 03 5 Package table 04 6 Picture and explanation for product 04 7 Explanation of button 05 8 Installation 05 9 Therapy 06 10...

Page 3: ...ified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation Por...

Page 4: ...al institutional use The YH 480A consists of main device water tank power adapter and cable Mask and heated tubing are purchased 3 Scope of application BreathCare PAP is a kind of the device which pro...

Page 5: ...t Caution Please change the distilled water in the water tank every day Do not add the distilled water above the maximum water level mark in case the water may enter the heated tubing and the device A...

Page 6: ...the highest heating level The heated tubing is used to heat the air so that the treatment will become more comfortable The heating level of the heated tubing 0 5 level equivalent to the following temp...

Page 7: ...Initial Pressure Therapy Pressure Pressure Unit FPS Level Ramp minutes CPAP 4 0 10 0 cmH2O 2 30 Clinical Menu Picture 5 1 Clinical menu CPAP Picture 5 2 Clinical menu APAP 10 2 1 Set Mode CPAP or APAP...

Page 8: ...e device has been disconnected from the power supply whether the power cord has been unplugged and whether the water tank of the device has cooled down Make sure the plate has cooled down to room temp...

Page 9: ...d fill it with distilled room temperature water up to the maximum water level mark Close the water tank and insert it into the side of the device Connect the heated tubing firmly to the air outlet loc...

Page 10: ...Press dial to turn on the screen light Check the connection of power ensure it connect with device firmly My water tank is leaking Water tank assemble incorrectly or it is broken Check the water asse...

Page 11: ...g Heated tubing Plastic hose about 1 8 m Maximum volume of water tank 260 10 mL Air outlet 22 mm complies with ISO 5356 1 2015 Temperature Maximum temperature of heater plate 55 4 131 14 4 Cut out 11...

Page 12: ...rning hot surface Class ii equipment Use by date Batch code Temperature limitation for storage and transport CN Country of manufacture china Date of manufacture is on the right side of the symbol or b...

Page 13: ...y test Compliance level Electrostatic discharge ESD IEC 61000 4 2 8kV contact 15kV air Electrical fast transient burst IEC 61000 4 4 2kV for power supply lines Surge IEC 61000 4 5 1kV differential mod...

Page 14: ...The 1 m test distance is permitted by IEC 61000 4 3 a For some services only the uplink frequencies are included b The carrier shall be modulated using a 50 duty cycle square wave signal c As an alter...

Page 15: ...any service organization that has not been expressly authorized by Yuwell to perform such repairs Any damage caused as accident Act of God or human factor Product which is not involved in Quality warr...

Page 16: ......

Reviews: