14.1 General troubleshooting:

Trouble

Cause

Solution

Air is leaking from 
around my mask.

Mask may be fitted incorrectly;

The size of the mask is not 
suitable.

Ensure your mask is fitted correctly;

See your mask user guide to check your 

mask fit and seal;

Please refer to the nose measurement 
card to choose the appropriate mask size.

I am getting a dry or 
blocked nose. 

Humidity level may set too low;
For the blocked nose:
History of nasal disease;
The pressure parameter may 
be set improperly.

Adjust the humidity level;
Use after treatment by a doctor;
Adjust the pressure parameter.

I am getting droplets 
of water on my nose, 
in the mask and 
heated tubing. 

The humidity level may be set 
too high;
Condensation forms in the 
piping mask due to cold 
weather.

Adjust the humidity level;
Open the heated tubing to solve 
condensation.

My mouth is very dry 
and uncomfortable. 

Air may be escape from your 
mouth;
The humidity level is not on or 
may be set too low.

You may need a chin strap to keep your 
mouth closed or a full face mask;
Increase the humidity level. 

Air pressure in my 
mask seems too 
high (it feels like I am 
getting too much air). 

Ramp may be turned off;
The pressure parameter may 
be set too high.

Enable your ramp option;
Decrease the pressure.

Air pressure in my 
mask seems too low 
(it feels like I am not 
getting enough air).

Ramp may be in  progress;
The pressure parameter may 
be set too low.

Start your therapy after the setting 
pressure reached or turn ramp time off;
Increase the pressure. 

My screen is black.

After therapy start, the screen 
backlight went off; 
In other case, the power do not 

connect firmly. 

Press dial to turn on the screen light; 
Check the connection of power, ensure it 

connect with device firmly.

My water tank is 
leaking.

Water tank assemble 
incorrectly or it is broken. 

Check the water assemble correctly; 
Contact your provider or Yuwell if your 
water tank is broken.

The heated tubing 
does not work.

The heated tubing is not 
properly connected;
The function of the heated 
tubing is not enabled;
The heated tubing is broken.

Checking the connection of the heated 
tubing, when the connection is successful, 
the icon   will be shown in the therapy 

page;
Enable the heated tubing function in the 
setting page;
Or contact your provider or Yuwell.

 

f

14.2 Other trouble: 

Message in LCD Meaning of error

Solution

ERROR 1

There is a mistake of pressure sensor

Please contact your provider or Yuwell

ERROR 2

There is a mistake of flow sensor 

Please repower your device or contact 
your provider or Yuwell

ERROR 3

There is a mistake of temperature 
sensor

Please contact your provider or Yuwell

ERROR 4

Functional safety detection is abnormal Please contact your provider or Yuwell

ERROR 5

There is a parameter out of the 
corresponding range

Please repower the device or contact 
your provider or Yuwell 

ERROR 6

The pressure is out of range

Replace the filter with a new one or 

repower the device or contact your 
provider or Yuwell 

ERROR 7

It is failed to supply power to the 

humidifier

Please contact your provider or Yuwell

ERROR 8

There is a mistake of blower

Please contact your provider or Yuwell

ERROR 9

There is a mistake of RTC ( loss power ) Please adjust the time of the device or 

contact your provider or Yuwell 

15. Specification of technology

Items

Specifies 

Power

Input :100-240VAC, 50-60Hz, 2Amax
Output:24V DC, 5A

Heated tubing

Power supply connection: 0-24VDC, 2A (max)
Connection port output: 0-24VDC, 2A (max)

Summary of Contents for YH-480A

Page 1: ...ay Pressure Units Suzhou Yuyue Medical Technology Co Ltd No 9 Jinfeng Road Suzhou Science Technology Town 215163 Suzhou Jiangsu PRC TEL 86 0512 67373001 web address www yuyue com cn IFU V 68 version 0...

Page 2: ...uction 03 3 Scope of application 03 4 Contraindications and adverse effects 03 5 Package table 04 6 Picture and explanation for product 04 7 Explanation of button 05 8 Installation 05 9 Therapy 06 10...

Page 3: ...ified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation Por...

Page 4: ...al institutional use The YH 480A consists of main device water tank power adapter and cable Mask and heated tubing are purchased 3 Scope of application BreathCare PAP is a kind of the device which pro...

Page 5: ...t Caution Please change the distilled water in the water tank every day Do not add the distilled water above the maximum water level mark in case the water may enter the heated tubing and the device A...

Page 6: ...the highest heating level The heated tubing is used to heat the air so that the treatment will become more comfortable The heating level of the heated tubing 0 5 level equivalent to the following temp...

Page 7: ...Initial Pressure Therapy Pressure Pressure Unit FPS Level Ramp minutes CPAP 4 0 10 0 cmH2O 2 30 Clinical Menu Picture 5 1 Clinical menu CPAP Picture 5 2 Clinical menu APAP 10 2 1 Set Mode CPAP or APAP...

Page 8: ...e device has been disconnected from the power supply whether the power cord has been unplugged and whether the water tank of the device has cooled down Make sure the plate has cooled down to room temp...

Page 9: ...d fill it with distilled room temperature water up to the maximum water level mark Close the water tank and insert it into the side of the device Connect the heated tubing firmly to the air outlet loc...

Page 10: ...Press dial to turn on the screen light Check the connection of power ensure it connect with device firmly My water tank is leaking Water tank assemble incorrectly or it is broken Check the water asse...

Page 11: ...g Heated tubing Plastic hose about 1 8 m Maximum volume of water tank 260 10 mL Air outlet 22 mm complies with ISO 5356 1 2015 Temperature Maximum temperature of heater plate 55 4 131 14 4 Cut out 11...

Page 12: ...rning hot surface Class ii equipment Use by date Batch code Temperature limitation for storage and transport CN Country of manufacture china Date of manufacture is on the right side of the symbol or b...

Page 13: ...y test Compliance level Electrostatic discharge ESD IEC 61000 4 2 8kV contact 15kV air Electrical fast transient burst IEC 61000 4 4 2kV for power supply lines Surge IEC 61000 4 5 1kV differential mod...

Page 14: ...The 1 m test distance is permitted by IEC 61000 4 3 a For some services only the uplink frequencies are included b The carrier shall be modulated using a 50 duty cycle square wave signal c As an alter...

Page 15: ...any service organization that has not been expressly authorized by Yuwell to perform such repairs Any damage caused as accident Act of God or human factor Product which is not involved in Quality warr...

Page 16: ......

Reviews: