28
Operation
Backup Data
→
Backup Data
This function allows you to backup logged data to a flash drive based on
Site, Data ID, and log date (
Figure 42
). A USB female to micro USB male
adapter is included with new instruments for this data backup.
NOTE:
The USB storage device must be formatted as FAT32, not
NTFS or exFAT. The handheld will only support FAT32.
If the box next to “
Include Sensor Info
” is checked, each data set will
be sent to a flash drive as a separate file with sensor serial number and
sensor software information included. If the box is not checked (default),
all data sets will be sent in a single backup file with no sensor serial
number or sensor software information.
NOTE:
It is suggested to send data to the USB flash drive as
a single file (i.e. box is not checked) unless this sensor
information is needed. This makes importing the data much
faster and easier.
Once the filter settings are configured, select
Backup Data
to send the
data to a flash drive. The data is exported in a CSV file.
If the data backup is not successful, ensure the correct filter criteria are
selected and the USB connection indicator can be seen at the top of the
screen (
Figure 9
).
Delete Calibration Record
→
Delete Calibration Record
To permanently delete the Calibration Record file from the instrument,
select
Yes
, then push the
ENTER
key (
Figure 44
).
Figure 42
Backup Data
Figure 43
Micro USB female
connector
Figure 44
Delete Calibration Record