APPENDIX A SPECIFICATIONS
Operating Environment
Medium: fresh, sea, or polluted water and most other liquid solutions.
Temperature: -5 to +65 °C
Depth: 0 to 10, 0 to 25, 0 to 50, or 0 to 100 feet (depending on cable length)
Storage Temperature:
-10 to +50 °C
Material:
ABS, Stainless Steel, and other materials
Dimensions:
Height:
9.5 inches
(24.13 cm)
Thickness:
2.2 inches
(5.6 cm)
Width:
3.5 inches max.
(8.89 cm)
Weight:
1.7 pounds (w/ 10’ cable)
(.77 kg)
Display:
2.3”W x 1.5”L
(5.8cm W x 3.8cm L)
Power:
9 VDC -6 AA-size Alkaline Batteries (included)
Approximately 100 hours operation from each new set of batteries
Water Tightness:
Meets or exceeds IP65 standards
Extensive testing of the YSI Model 85 indicates the following typical performance:
Measurement
Range
Resolution
Accuracy
Conductivity
0 to 499.9
µ
S/cm
0.1
µ
S/cm
±
.5% FS
0 to 4999
µ
S/cm
1.0
µ
S/cm
±
.5% FS
0 to 49.99 mS/cm
.01 mS/cm
±
.5% FS
0 to 200.0 mS/cm
0.1 mS/cm
±
.5% FS
Salinity
0 to 80 ppt
.1 ppt
±
2%, or
±
0.1 ppt
Temperature
-5 to +65 °C
0.1 °C
±
0.1 °C (
±
1 lsd)
Dissolved Oxygen
0 to 200 % Air Sat.
0.1% Air Saturation
±
2% Air Saturation
0 to 20 mg/L
0.01 mg/L
±
0.3 mg/L
Adjustable Conductivity Reference
Temperature
: 15°C to 25°C
Adjustable Temperature Compensation Factor for Conductivity:
0% to 4%
Temperature Compensation:
Automatic
Range:
Autoranging for Dissolved Oxygen
User selected or Autoranging for Conductivity
Summary of Contents for 85
Page 6: ...Introduction Section 1 YSI Incorporated Model 85 2 ...
Page 10: ...Preparing the Meter Section 2 YSI Incorporated Model 85 6 ...
Page 12: ...Preparing the Probe Section 3 YSI Incorporated Model 85 8 ...
Page 14: ...Overview of Operation Section 4 YSI Incorporated Model 85 10 ...
Page 18: ...Calibration Section 5 YSI Incorporated Model 85 14 ...
Page 23: ...Making Measurements Section 7 YSI Incorporated Model 85 19 ...
Page 24: ...Making Measurements Section 7 YSI Incorporated Model 85 20 ...
Page 28: ...Maintenance Section 9 YSI Incorporated Model 85 24 ...
Page 36: ...Discussion of Measurement Errors Section 11 YSI Incorporated Model 85 32 ...
Page 43: ...Warranty and Repair Section 13 YSI Incorporated Model 85 39 ...
Page 45: ...Accessories and Replacements Section 14 YSI Incorporated Model 85 41 ...
Page 47: ...Specifications Appendix A YSI Incorporated Model 85 43 ...
Page 53: ...Required Notice Appendix C YSI Incorporated Model 85 49 ...
Page 55: ...Conversion Chart Appendix D YSI Incorporated Model 85 51 ...
Page 59: ...Calibration Values Table Appendix F YSI Incorporated Model 85 55 ...