SECTION 9 MAINTENANCE
9.1 CLEANING AND STORAGE
The single most important requirement for accurate and reproducible results in conductivity
measurement is a clean cell. A dirty cell will change the conductivity of a solution by contaminating
it.
NOTE:
ALWAYS RINSE THE CONDUCTIVITY CELL WITH CLEAN WATER AFTER
EACH USE.
To clean the conductivity cell:
1.
Dip the cell in cleaning solution and agitate for two to three minutes. Any one of the foaming
acid tile cleaners, such as Dow Chemical Bathroom Cleaner, will clean the cell adequately.
When a stronger cleaning preparation is required, use a solution of 1:1 isopropyl alcohol and 1
N HCl. Remove the cell from the cleaning solution.
2.
Use the nylon brush (supplied) to dislodge any contaminants from inside the electrode chamber.
3.
Repeat steps one and two until the cell is completely clean. Rinse the cell thoroughly in
deionized, or clean tap water.
4.
Store the conductivity cell in the meter storage chamber.
NOTE: See Section 11, Dissolved Oxygen Probe Precautions for instructions on cleaning the
dissolved oxygen electrodes.
Summary of Contents for 85
Page 6: ...Introduction Section 1 YSI Incorporated Model 85 2 ...
Page 10: ...Preparing the Meter Section 2 YSI Incorporated Model 85 6 ...
Page 12: ...Preparing the Probe Section 3 YSI Incorporated Model 85 8 ...
Page 14: ...Overview of Operation Section 4 YSI Incorporated Model 85 10 ...
Page 18: ...Calibration Section 5 YSI Incorporated Model 85 14 ...
Page 23: ...Making Measurements Section 7 YSI Incorporated Model 85 19 ...
Page 24: ...Making Measurements Section 7 YSI Incorporated Model 85 20 ...
Page 28: ...Maintenance Section 9 YSI Incorporated Model 85 24 ...
Page 36: ...Discussion of Measurement Errors Section 11 YSI Incorporated Model 85 32 ...
Page 43: ...Warranty and Repair Section 13 YSI Incorporated Model 85 39 ...
Page 45: ...Accessories and Replacements Section 14 YSI Incorporated Model 85 41 ...
Page 47: ...Specifications Appendix A YSI Incorporated Model 85 43 ...
Page 53: ...Required Notice Appendix C YSI Incorporated Model 85 49 ...
Page 55: ...Conversion Chart Appendix D YSI Incorporated Model 85 51 ...
Page 59: ...Calibration Values Table Appendix F YSI Incorporated Model 85 55 ...