SECTION 3 PREPARING THE PROBE
The YSI Model 85 dissolved oxygen probe is shipped dry. The protective membrane cap on
the probe tip must be removed and replaced with KCl solution and a new membrane cap
before using the probe. Follow the instructions below to install KCl solution and the new
membrane cap.
3.1 MEMBRANE CAP INSTALLATION
To install a new membrane on your YSI Model 85 dissolved oxygen probe:
1.
Unscrew and remove the probe sensor guard.
2.
Unscrew and remove the old membrane cap.
3.
Thoroughly rinse the sensor tip with distilled water.
4.
Prepare the electrolyte according to the directions on the KCl solution bottle.
5.
Hold the membrane cap and fill it at least 1/2 full with the electrolyte solution.
6.
Screw the membrane cap onto the probe moderately tight. A small amount of
electrolyte should overflow.
7.
Screw the probe sensor guard on moderately tight.
Screw Cap on
moderately tight
Unscrew Guard
Unscrew Cap
Fill Membrane
with KCL Solution
Screw Guard on
moderately tight
Summary of Contents for 85
Page 6: ...Introduction Section 1 YSI Incorporated Model 85 2 ...
Page 10: ...Preparing the Meter Section 2 YSI Incorporated Model 85 6 ...
Page 12: ...Preparing the Probe Section 3 YSI Incorporated Model 85 8 ...
Page 14: ...Overview of Operation Section 4 YSI Incorporated Model 85 10 ...
Page 18: ...Calibration Section 5 YSI Incorporated Model 85 14 ...
Page 23: ...Making Measurements Section 7 YSI Incorporated Model 85 19 ...
Page 24: ...Making Measurements Section 7 YSI Incorporated Model 85 20 ...
Page 28: ...Maintenance Section 9 YSI Incorporated Model 85 24 ...
Page 36: ...Discussion of Measurement Errors Section 11 YSI Incorporated Model 85 32 ...
Page 43: ...Warranty and Repair Section 13 YSI Incorporated Model 85 39 ...
Page 45: ...Accessories and Replacements Section 14 YSI Incorporated Model 85 41 ...
Page 47: ...Specifications Appendix A YSI Incorporated Model 85 43 ...
Page 53: ...Required Notice Appendix C YSI Incorporated Model 85 49 ...
Page 55: ...Conversion Chart Appendix D YSI Incorporated Model 85 51 ...
Page 59: ...Calibration Values Table Appendix F YSI Incorporated Model 85 55 ...