YSI Incorporated
Model 5000/5100 Service Manual
12
8.15
mg/L
Calibrate
0.0
ppt
740
mmHg
24.38°C 11:33
AM
AUTO CAL
DO CAL
SETUP
DIAGNOSTICS
7. Press the [DO CAL] soft-key to enter the manual DO calibration menu.
98.0
% 8.15
mg/L
Cal %
0.0
ppt
740
mmHg
24.38°C 11:33
AM
UP
DOWN
DIGIT
NEXT
8. Verify that the salinity is set to the salinity value of the calibration environment (0.0 ppt for
air calibration). ). See
Calibration, Salinity
.
Make sure that the display readings are stable, then enter the calibration value in percent
saturation, using the [UP], [DOWN] and [DIGIT] soft-keys.
If you wish to calibrate in milligrams per liter (instead of percent), use the [NEXT] soft-key to
select mg/L, then enter the calibration value in milligrams per liter as shown below.
98.0
*
% 08.15
mg/L
Cal mg/L
0.0
ppt
740
mmHg
24.38°C 11:33
AM
UP
DOWN
DIGIT
NEXT
9. Press [ENTER] to confirm your calibration. The screen will momentarily display
“D.O.
CALIBRATION SAVED”
as shown below.
98.0
*
% 8.15
mg/L
Calibration
0.0
ppt
740
mmHg
24.38°C 11:33
AM
D. O. CALIBRATION SAVED
UP
DOWN
DIGIT
NEXT
Press [MODE] to return to the Main mode. The instrument is now calibrated and ready to
measure dissolved oxygen and temperature.
Flashing digit
Flashing digit
Message