YSI Incorporated
Model 5000/5100 Service Manual
18
Code/Problem Error
Possible
Cause
Correction
E4:
Check
Membrane
DO High Current
Membrane too thin
Install new membrane and
KCl
Electrodes need resurfacing
See probe maintenance
instructions
Short circuit in probe, cable or
connector
Replace defective assembly
E5:
Check
Membrane
DO unstable
Membrane damaged
Replace membrane and KCl
Probe electrodes needs
servicing
See probe maintenance
instructions
Connector
improperly
installed
Check probe connection
Intermittent connection in
cable or plug
Replace defective assy
Probe malfunction
Replace defective assy