No battery
The “No battery” alarm appears if you remove the alkaline battery from the battery
compartment for more than 5 minutes while the insulin pump is in run mode. Confirm
by tapping and insert a new AAA alkaline battery (LR03). Follow the procedure
on page 53
of this Caregiver Guide. The “No battery” alarm only appears if the battery
was removed while the insulin pump was in run mode. If the battery is removed in
stop mode, the insulin pump adopts storage status after 5 minutes.
Battery empty
The “Battery empty” alarm appears if the battery no longer has sufficient power and
has to be changed. Confirm by tapping and insert a new AAA alkaline battery (LR03).
Follow the procedure on page 53
of this Caregiver Guide.