4/6:
Confirm that you have disconnected
the infusion set from the body by remov-
ing the tubing cap from the cannula base
and confirm by tapping . The mylife
™
YpsoPump
®
vibrates briefly, the threaded
rod advances up to the plunger of the
cartridge and the infusion set is primed with
the set amount of insulin.
5/6:
During the priming procedure, keep
the insulin pump in an upright position
with the adapter facing upwards and tap
the insulin pump lightly on your open
palm to remove the air bubbles from the
infusion set. The screen counts the
delivered priming volume upwards until
the set value has been reached. You can
cancel the priming procedure at any time
by tapping .
6/6:
Repeat the priming procedure with the
appropriate priming volume until there are
no more air bubbles in the cartridge, in the
adapter or in the tubing and until insulin
has emerged from the end of the tubing of
the infusion set.
If the temporary basal rate function or a bolus is activated, the prime infusion set
function is not available. The prime infusion set function also cannot be executed if
the cartridge is empty.