How to change the cartridge and the infusion set
1/8:
Always disconnect the tubing cap
of the infusion set from the body first.
Then carefully remove the tape from the
body.
Removing the cartridge and the infusion set
2/8:
Open the main menu and tap the
“Cartridge change and current cartridge
level” icon.
3/8:
Tap the “Return threaded rod” icon.
4/8:
Confirm by tapping . The mylife
™
YpsoPump
®
vibrates briefly.
5/8:
The threaded rod returns and the
percentage is reduced to 0 %.
6/8:
Tap to start and complete the
self-test.
For the mylife
™
YpsoPump
®
, you may only use a self-filled 1.6-ml reservoir
(mylife
™
YpsoPump
®
Reservoir) compatible with the mylife
™
YpsoPump
®
.