Page 8
91000-90(B)
SET10nnnn
Set output interval. Sets the time interval
between measurements in one millisecond
increments. Lower values increase power
consumption when continuous measurements
are taken. Default is 1000.
Where nnnn=
0000-9999
Usage example:
>SET101000
Sets output interval to 1000 milliseconds
(1 second) (Note: If output interval is set
too low, the
ResponseONE
will automatically
calculate the minimum output interval.)
SET12nnn
Set sample count. The rate at which the sensor
internally takes complete sonic wind samples
is greater than 200 times per second. This
command sets the number of internal samples
used to calculate the median measurement
result. Default is 16.
Where nnn=
000-800
Usage example:
>SET12016
Sets sample count to 16.
More internal samples consume more power
while providing greater immunity to conditions
like turbulent high-speed wind. Fewer samples
consume less power while providing less
immunity to disruptive conditions. This command
interacts with the OUTPUT INTERVAL setting.
More samples may require a longer output
interval.
SET13n
Set ASCII output format. Determines whether
ASCII and ASCII POLLED serial outputs provide
wind data in either Polar (speed and direction)
or Cartesian (UV) form. Default is Polar.
Where n= 0
Polar
1
Cartesian (UV)
Usage example:
>SET130
Sets output format to polar.
SET25n
Turn compass on or off. Default is 0.
Where n= 0
Off
1
On
Usage example:
>SET251
Turns compass on. Note: This command should
only be used if the sensor has been calibrated at
current location. Calibration automatically turns
on compass.
SET26nddmm
Set compass (magnetic) declination.
Magnetic declination is the angle
between magnetic north and true north.
Where
n=
E, W, or 0 (East, West, or none)
dd=
00 to 90 degrees
mm=
00 to 60 minutes
Usage example:
>SET26W0606
Set declination to West 6° 6’ (zip code 49686)
SET77
Save all settings to non-volatile memory. Any
settings which have not been saved will be
lost when power is removed. Saved settings
are loaded at power up. This command may
be used any time the sensor is in COMMAND
MODE. Note:
ResponseONE
CONFIG
program automatically sends this command
when <Send Parameters> button is pressed.
XX
Returns the sensor to OPERATE MODE
RPTV
Report current settings
??
Display command list
HELP
Display command list