50
WARNING
:
1
.
The product must be away from machine with high current or
strong magnetic.
2
.
Do not immerse the probe in any type of liquid or detergent.
3
.
Disinfection of probe with gas or heating is strictly
prohibited.
4
.
It is strictly prohibited to immerse probe socket in any
solution. Probe surface is waterproof but only probe head.
Please be careful while using coupling agent and cleaning or
disinfecting probe surface.
5
.
Users are not allowed to repair without authorization in case
of probe failure, Product must be sent back to the company for
repair.
Check the following items after cleaning and disinfection
:
Inspection
items
Inspection methods
Acceptance Criteria
Acoustical lens Visual inspection of lens surface and connection
between lens and shell of the acoustic head
No damages or cracks on lens
surface;
No cracks at connection between lens
and shell.
Shell
Visual inspection of shell surface
No damages or cracks on shell
surface
Shell
connection
Visual inspection of shell connection
No cracks or lack of gluing at joints
of the shell
Residue
Visual inspection of probe surfaces, especially
pits and narrow areas
No residual detergent or disinfectant