47
7.2
Probe cover
In order to reduce the spread of the disease, it is necessary to take some protective measures. In
clinic, we always use ultrasound probe cover to prevent infections.
. It is strongly recommended that a qualified and bio-clean probe cover should be used when
carrying out cavity inspection.
The probe has to wear a probe protection cover or before the probe is used to examine the
human body. Please use the probe protection sheath accepted by the market.
CAR
EFUL
:
1.
In order to avoid infection, the probe cover or sheath can be used
only one time, and broken ones should never be used.
2.
The probe sheath, made of natural latex and talc, might cause
allergic reactions to some people.
3.
Do not use expired probe covers. Check the expiration date of the
probe covers before using it.
Operation steps (for reference only)
:
1. Put coupling agent into the probe cover to get clear images
2.Ensure sterility when the probe goes into the protective cover. Making the protective cover
properly tense to avoid any wrinkles and bubbles, do not over pull the probe cover.
3. To fasten the probe cover with a string for safety concern.
4. Check the protective covers and ensure no broken ones.
7.3
Check up and maintenance
7.3.1 Check up
Frequently check the sockets. If any parts found damaged, repair or replace it immediately.
Check acoustic lens regularly. If any bubbles found, repair or replace the broken parts right
away.
7.3.2 Service life
According to the manufacturer's design, production and other related documents, the service
life of this probe is about 5 years. The materials which form the probe may gradually age as its
running time goes on. If the product continues to be used beyond its service life, its performance
will decline and the failure rate will increase obviously.