8 ELECTROMAGNETIC
COMPATIBILITY
8.1 General
The HFG02 has been designed to meet the requirements of EU Directive
2014/30/EU. Compliance was demonstrated by testing to the requirements of the
following standards: -
8.2 Emissions
EN61326-1 (2013) consisting of
EN55011 (2009) + A1 (2010) Radiated RF emissions (Enclosure) and
Conducted RF emissions (applies to the HFG02 supply adapter), Class A
8.3 Immunity
Enclosure:
EN61326-1 (2013) consisting of
EN61000-4-2 (2009) Electrostatic Discharge immunity, 4 kV Air & Contact
Discharge.
EN61000-4-3 (2006) + A1 (2008) + A2 (2010) Radiated RF immunity
(Enclosure), 3 V/m.
AC power:
These apply to the HFG02 supply adapter only:
EN61326-1 (2013) consisting of
EN61000-4-11 (2004) Voltage dips/short interruptions
EN61000-4-4 (2004) + A1 (2010) Fast Transient common mode (AC mains
port)
EN61000-4-5 (2006) Surge (AC mains port)
EN61000-4-6 (2009) Conducted RF immunity (AC mains port)
EN61000-3-2 (2006) Harmonic emission (AC mains port)
EN61000-3-3 (2013) Voltage fluctuation (AC mains port)
8.4 Non-Applicability
EN61326-1 (2013) consisting of
EN61000-4-4 (2004) Fast Transient common mode (signal ports)
EN61000-4-5 (2006) Surge (signal ports)
EN61000-4-6 (2009) Conducted RF immunity (signal ports)
These phenomena are deemed non applicable on the basis that the HFG02 has
no relevant signal ports. The AC test port of the HFG02 may be looked upon as a
signal port but the captive lead is less than 3m in length hence the tests do not
apply.
The test port is not connected directly to the public low voltage supply network and
so tests relating to AC ports are only applicable to the AC/DC adapter used for the
HFG02 supply input.