HFG02 Operation Manual
York EMC Services
Page 17 of 25 Doc. No. Y0321/DOC/024 Rev. 3
6 O
UTPUT
P
ERFORMANCE
Please refer to the test results supplied with the unit for its unique performance.
Typical performance results are shown in the sections below.
6.1 Flicker Load Operation
The typical flicker disturbance generated by the HFG02 in each of the steady flicker
modes when tested according to EN/IEC 61000-3-3 is shown in Table 3.
Flicker Mode
Setting
Load Operation
Pst
Plt
8.3/8.6 Hz
8.33 Hz flicker (50 Hz Test Supply detected)
8.6 Hz flicker (60 Hz Test Supply detected)
1.10
1.06
1.10
1.06
1 Hz
1 Hz flicker
(50 Hz or 60 Hz Test Supply detected)
0.45
0.45
Table 3: Typical Pst results measured to EN/IEC 61000-3-3
6.2 Harmonics Load Operation
The typical current harmonics disturbance generated by the HFG02 in each of the
harmonics modes when tested according to EN/IEC 61000-3-2 is shown below:
Figure 7: Disturbances generated in Fixed Harmonic mode