Inaccurate
measurement
Waiting time is not enough. Wait at last 30 minutes.
The measurement method
is not correct.
Use the wrong
measurement mode.
According to the requirements of
the user manual to measuring.
The surface of the probe tip
or forehead cap is dirty.
According to the requirements of
the user manual to cleaning.
In forehead measure
mode, the measuring
distance is not correct .
Please respect the measuring
distance shall be within 1.0cm to
3.0cm.
In ear measure mode, the
probe tip isn’t aimed at the
eardrum and/or isn’t
closely matched with the
ear canal
Ensure that the ear canal is clean
and the probe tip and ear canal are
closely matched when measuring
ear temperature.
The Bluetooth
connection failed
The Bluetooth module of
the thermometer failed to
initialize
Remove the battery from the
thermometer and reinstall it to
reinitialize
The distance between the
thermometer and the
Bluetooth APP is too large
Make sure the distance between
thermometer and Bluetooth APP is
within 10 meters
Remark:
If the above methods still do not solve the problem, please contact the
manufacturer, contact information see the last page.
16. Standard List
Shenzhen Yolanda Technology Co., Ltd. declares that the TH10A complies with following
applicable standards:
IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ASTM E 1965-98, ISO 10993-1, ISO 10993-
5, ISO 10993-10 and IEC 62304, ISO 1497, ISO 80601-2-56, EN 1401, ISO 15223-1.
17. Disposal
Dispose of the device in accordance with the regulations applicable at the
place of operation.