If Essential Performance is lost or degraded due to electromagnetic disturbances, this may result in
inaccurate measurement and lead to mislead patients, please read below important information before
to avoid possible electromagnetic disturbances.
Warning:
Using cell phone or microwave oven, HF surgical equipment, magnetic resonance imaging or other radio
radiant equipment near this product may cause malfunction or lead to loss of essential performance, which
means that the measurement accuracy will be affected.
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result
in improper operation. If such use is necessary, this equipment and the other equipment should be observed
to verify that they are operating normally.
Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the monitor. Otherwise,
degradation of the performance of this equipment could result.
Use of accessories, transducers other than those specified or provided by the manufacturer of this
equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of
this equipment and result in improper operation.
Caution:
Security, antitheft, and radiofrequency identification (RFID) devices. Some electromagnetic anti-theft
systems and metal detectors such as those used at entrances or exits of department stores, libraries, and
other public places, and airport security screening devices may affect the monitor. Additionally, RFID
devices, which are often used to read identification badges, as well as some tag deactivation devices, such
as those used at payment counters at stores and loan desks at libraries, may also affect the monitor. Please
do not use monitor near these places. If you have to go through one of these devices, turn off your monitor.
Before each usage, checking the status of your monitor to ensure it can operating normally.
Using short-wave diathermy, microwave diathermy, or therapeutic ultrasound diathermy (all now referred to as
diathermy) and electrocautery devices near this product may cause malfunction or lead to loss of essential
performance, please do not use monitor near these equipment. Before each usage, observing the device to
verify that they are operating normally.
A list of cables and maximum length of cables is as follows:
Cables name
Cable length
Whether shielding
USB line
1m
No
Guidance and manufacture’s declaration – electromagnetic emission
The upper arm blood pressure monitor is intended for use in the electromagnetic environment specified below.
The customer of the user of the upper arm blood pressure monitor should assure that it is used in such an
environment.
Emission test
Compliance
Electromagnetic environment – guidance
Conducted and Radiated RF
emissions
CISPR 11
Group 1
Class B
The upper arm blood pressure monitor uses RF energy
only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference