6-5
IM 01B04J01-02E
Maintenance
1
2
3
App
1
2
3
5
6
7
8
4
DC Voltage/Current
Standard
Yokogawa GS200 or the equivalent
Power Supply
Digital Voltmeter
Yokogawa DM7560 or the equivalent
+
+
+
–
–
–
Digital Voltmeter
250
Ω
± 0.05%
ZERO
SPAN
5
3
1
2 4 6
8
7
J
K
D
B
H
F
C
A
F0601.ai
Decade Resistance
Boxes
Yokogawa 279301 or the equivalent
Power Supply
B
B
A
Digtal Voltmeter
Yokogawa DM7560 or the equivalent
Digital Voltmeter
250
Ω
± 0.05%
1 3 5
6
4
2
7
8
J
K
B D H
A C F
SPAN
ZERO
+
+
–
–
F0602.ai
Figure 6-1 mV DC Input Converter Adjustment
Figure 6-2 RTD Input Converter Adjustment
Note: The broken line shows
4 to 20 mA DC output
test circuit.
Note: The broken line shows
4 to 20 mA DC output
test circuit.
Note: The broken line shows
4 to 20 mA DC output
test circuit.
Note: The broken line shows
4 to 20 mA DC output
test circuit.
DC Voltage/Current
Standard
Yokogawa GS200 or the equivalent
Power Supply
r
Reference Junction Block
r: Resistance equivalent to that of
the actual lead wire.
Digital Voltmeter
Yokogawa DM7560 or the equivalent
Digital Voltmeter
250
Ω
± 0.05%
1 3 5
6
4
2
7
ZERO
SPAN
+
+
+
–
–
–
J
K
H
D
B
A C F
F0603.ai
Decade Resistance
Boxes
Decade Resistance
Boxes
Yokogawa 279301 or the equivalent
Power Supply
100%
0%
Digtal Voltmeter
Yokogawa DM7560 or the equivalent
Digital Voltmeter
250
Ω
± 0.05%
1 3 5
6
4
2
7
8
F
C
A
H
D
B
K
J
SPAN
ZERO
+
–
+
–
F0604.ai
Figure 6-3 Thermocouple Input Converter Adjustment
Figure 6-4 DC Potentiometer Input Converter
Adjustment
6.3 Adjustment