[:INPut]:FILTer?
Function
Queries all
input filter settings.
Syntax
[:INPut]:FILTer?
[:INPut]:FILTer:LINE
Function
Sets or queries the
line filter.
Syntax
[:INPut]:FILTer:LINE {<Boolean>}
[:INPut]:FILTer:LINE?
Example
:INPUT:FILTER:LINE OFF
:INPUT:FILTER:LINE?
-> :INPUT:FILTER:LINE 0
[:INPut]:FILTer:FREQuency
Function
Sets or queries the
frequency filter.
Syntax
[:INPut]:FILTer:FREQuency
{<Boolean>}
[:INPut]:FILTer:FREQuency?
Example
:INPUT:FILTER:FREQUENCY OFF
:INPUT:FILTER:FREQUENCY?
-> :INPUT:FILTER:FREQUENCY 0
[:INPut]:POVer?
Function
Queries the peak over-range information.
Syntax
[:INPut]:POVer?
Example
:INPUT:POVER? -> 0
Description • The peak over-range information of each
element is mapped as shown below. For the
response, the sum of the values of each bit is
returned in decimal format.
• For example, a response of 16 indicates that a
peak over-range is occurring at U3.
U1
I2 U2 I1
U3
I3
7
6
5
4
3
2
1
0
[:INPut]:CRANge?
Function
Sets or queries the
check range status.
Syntax
[:INPut]:CRANge?
Example
:INPUT:CRANGE? -> 0
Description • The CHECK RANGE LED status is mapped
as shown below. For the response, the sum
of the values of each bit is returned in decimal
format.
• For example, a response of 0 indicates
appropriate range. A response of 64 indicates
that a current over-range is occurring.
VL
VO VH
VP
AL
AH
AO
AP
7
6
5
4
3
2
1
0
VL : The voltage is at the condition for
reducing the auto range or less.
VH : The voltage exceeds the condition for
raising the auto range.
VO : The voltage is over-range.
VP : The voltage is peak over-range.
AL : The current is at the condition for
reducing the auto range or less.
AH : The current exceeds the condition for
raising the auto range.
AO : The current is over-range.
AP : The current is peak over-range.
• For information on the color of the LED and
the condition to light, see “Auto Range Monitor
Indications” in section 1.5 in the Getting
Started Guide, IM WT310E-02EN.