2-6
IM WT310E-17EN
2.4
Configuring the GP-IB Settings of This
Instrument
This section explains the following setting for controlling this instrument remotely through a GP-IB
interface:
Procedure
Follow the procedure indicated by the thick lines in the following menu.
Setting the GP-IB Address
2.
Select the
communication function.
1.
3
.
Set the address.
(Display D)
6
.
Press and to set the number.
Press to move the digit.
4.
5.
Set, close menu.
(Display B)
Note
• Only use one communication interface: USB, GP-IB, or Ethernet. If you send commands simultaneously
from more than one communication interface, this instrument will not execute the commands properly.
• When the controller is communicating with this instrument or with other devices through GP-IB, do not
change the address.
• Each device that is connected by GP-IB has its own unique address in the GP-IB system. This address
is used to distinguish between different devices. Therefore, you must assign a unique address to this
instrument when connecting it to a PC or other device.