3-4
IM 707235-01E
3.3
Self Test
If you believe that the module is not operating correctly, perform the self test according to
the following procedure with nothing connected to the input terminals of the module
being tested and the external trigger input terminal/external I/O connector of the
measuring station.
Executing Self Test
1. Select
Self Test
from the
System
menu of the WE7000 Control Software.
2. In the Self Test dialog box that appears, select the station name and enter the slot
number corresponding to the module, and click
Execute
. “Executing...” is displayed in
the Result display box.
→
Verifying Test Results
If a value other than 0 is displayed in the Result display box of the Self Test dialog box,
the module is probably malfunctioning. Please contact your nearest YOKOGAWA dealer
as listed on the back cover of this manual for repairs.