3-1
IM 707235-01E
T
roubleshooting and Maintenance
3
Chapter 3 Troubleshooting and Maintenance
3.1
Troubleshooting
• If servicing is necessary, or if the analyzer is not operating correctly after performing
the following corrective actions, contact your nearest YOKOGAWA dealer as listed on
the back cover of this manual.
• To verify that the module is operating correctly, perform the self test as described on
the next page.
Problem
Probable Cause/Corrective Action
Reference Page
Module does not operate.
Check to see that the module is installed correctly into the station.
2-1, *
Also, install the module into another slot, and check whether it will
operate there. If it operates in the other slot, the measuring station
is likely to have malfunctioned.
If the module is installed correctly and does not operate, the
connector might be bad or the IC may have malfunctioned. In either
case, contact your nearest YOKOGAWA dealer to have it repaired.
Waveform data cannot be
Check to see that the input cables are connected properly to the
2-3 to 2-5
acquired.
BNC terminals.
Noise enters the input signal.
If the signal line and the AC power supply line are close to each
—
other, move them apart.
Also make sure that the signal line is away from the noise source.
Change to a shielded signal cable if you are not already using one.
If the frequency of the input signal is known, remove the frequency
1-4
region above the frequency component of the input signal by setting
an input filter.
Measured values are not correct.
Check whether the ambient temperature and humidity are within
4-5
the allowed ranges.
If you did not allow a warm-up time of 30 minutes, try measuring
again after the warm-up time has passed.
Trigger does not activate
Check whether the trigger setting is adequate for the input source
1-2, 1-7
in the trigger operation panel.
If you are using the bus trigger signal, verify that the settings are
*
correct in the trigger source/time base source/arming setting dialog
box of the WE7000 Control Software.
The waveform monitor does
Check to see that the waveform monitor ON/OFF button, located to
1-2
not appear.
the right of the Start button of the operation, is not set to OFF.
* See the WE7000 User’s Manual (IM 707001-01E).