App-4
IM 704310-12E
Error in communication Query (400 to 499)
Code
Message
Action
Reference Page
410
Query INTERRUPTED
The transmission of the response was aborted.
2-2
Check the transmission/reception order.
420
Query UNTERMINATED
No response to be transmitted.
2-2
Check the transmission/reception order.
430
Query DEADLOCKED
Transmission/reception deadlock occurred.
2-2
Abort the transmission.
Limit the length of the program messages to 1024 bytes
including <PMT>.
440
Query UNTERMINATED after
The request order is not correct.
indefinite response
Do not enter a query after the
*IDN?
and
*OPT?
commands.
Error in Execution (600 to 899)
See User’s Manual IM704310-01E.
Error in System Operation (912 to 915)
Code
Message
Action
Reference Page
912
Fatal error in Communication-
Communication driver error occurred.
driver
Servicing is required.
914
A communication time-out has
Communication timeout error occurred.
occurred
Make sure to receive the data within the timeout time.
The communication line may be malfunctioning.
Other (350)
Code
Message
Action
Reference Page
350
Queue overflow
Read out the error queue.
4-5
Note
Code 350 indicates overflow of error queue. This code is returned as a response to the
"STATus:ERRor?"
query; it does not appear on the screen.