3-40
IM 704310-12E
3.11 MEMory Group
The commands in this group deal with the transmission of the measured data to external devices.
:MEMor y
:
FORMat
<Space>
ASCii
BINar y
?
BYTeorder
<Space>
LSBFirst
MSBFirst
?
DATaselect
<Space>
FREQuency
MEASuredata
TSTamp
?
STARt
<Space>
<Time>
<NRf>
?
END
<Space>
<NRf>
?
SIZE
?
SEND
?
<Time>
;
BLOCk
<Space>
?
<NRf>
?
CLEar
MEMory?
Function
Queries all settings related to the transmission
of the measured data to external devices.
Syntax
:MEMory?
Example
:MEMORY?
→
:MEMORY:DATASELECT FREQUENCY;
FORMAT ASCII;BYTEORDER LSBFIRST;
START 0.000E-09;END 1.000E-09;BLOCK 1
MEMory:BLOCk
Function
Sets the block during block sampling or queries
the current setting.
Syntax
:MEMory:BLOCk <NRf>
:MEMory:BLOCk?
Example
:MEMORY:BLOCK 1
:MEMORY:BLOCK?
→
:MEMORY:BLOCK 1
MEMory:BYTeorder
Function
Sets the transmission byte order of the binary
data or queries the current setting.
Syntax
:MEMory:BYTeorder {LSBFirst|MSBFirst}
:MEMory:BYTeorder?
Example
:MEMORY:BYTEORDER LSBFIRST
:MEMORY:BYTEORDER?
→
:MEMORY:BYTEORDER LSBFIRST
3.11 MEMory Group
MEMory:CLEar
Function
Clears the measured data.
Syntax
:MEMory:CLEar
Example
MEMORY:CLEAR
MEMory:DATaselect
Function
Sets the data to transmit or queries the current
setting.
Syntax
:MEMory:DATaselect {TSTamp|MEASuredata|
FREQuency}
:MEMory:DATaselect?
Example
:MEMORY:DATASELECT TSTAMP
:MEMORY:DATASELECT?
→
:MEMORY:DATASELECT TSTAMP
Description
If time stamp (
TSTamp
) is specified when the
measurement mode is set to histogram or
frequency (
FREQuency
) is specified when the
mode is set to time stamp, an error occurs
when executing the
MEMory:SEND?
command.