Overview of the GP-IB Interface
1-3
IM 704310-12E
1
1.3
GP-IB Interface Functions and Specifications
GP-IB Interface Functions
Listener function
• Allows you to make the same settings which you can make using the panel keys on
the instrument (except for the power ON/OFF and GP-IB communications settings).
• Receives commands from a controller requesting output of set-up and waveform data.
• Also receives status report commands.
Talker function
Outputs set-up and waveform data.
Note
The listen-only, talk-only, and controller functions are not available on this instrument.
Switching between Remote and Local Modes
Switching from Local to Remote Mode
Remote mode is activated when a REN (Remote Enable) message is received from a
controller while local mode is active.
• The word [Remote] appears in the top right corner of the screen (see page 1-1).
• All front panel keys except the LOCAL key are now inoperative.
• Settings that were entered in local mode are retained.
Switching from Remote to Local Mode
Pressing the Local key in remote mode puts the instrument in local mode. However, this
is not possible if Local Lockout has been set by the controller (page 1-7).
• The word [Remote] in the top right corner of the screen disappears (see page 1-1).
• All front panel keys are operative.
• Settings that were entered in remote mode are retained.
1.3 GP-IB Interface Functions and Specifications