1 PREFACE

1.1 Introduction

The SC24V sensor with Variopin connector is 
not a conventional pH sensor, but a differential 
sensor that uses two dissimilar ion sensitive 
glass membranes to generate the pH signal. 
The measuring element is the same as a 
conventional pH electrode, and the reference 
similar to a Sodium ion electrode. Therefore the 
SC24V can only be used in combination with 
pH analyzers that feature:
•  Dual high impedance sensor inputs
•  Adjustable setting for the Isothermal Point
•  Adjustable temperature coefficient

The SC24V differential measuring principle 
has significant advantages over conventional 
electrode designs in a variety of difficult 
applications. The hermetically sealed reference 
eliminates exposure of the reference fill solution 
to the process, as there is no reference junction 
to be affected by aging, plugging or poisoning 
from the process. This greatly reduces 
maintenance problems since the reference 
stability of the differential sensor is then only 
dependant on the sodium concentration of the 
process.

The SC24V is ideally suited for difficult 
applications where conventional electrodes 
suffer from reference problems, especially 
when high temperatures and high pressures 
are involved. 70-80% of industrial users will fully 
benefit from using the SC24V, as they will have 
sufficient sodium levels in their processes for a 
stable measurement. Examples of applications 
where the sensor has proven to be successful 
are Chlorine manufacturing by electrolysis of 
brine and Bio-ethanol fermentation.
The only applications where the SC24V cannot 
be used are those where the salt concentration 

is fluctuating more than 30% or is too low, such 
as in pure and ultrapure water processes.

1.2 Unpacking and Checking

Upon delivery, unpack the sensor carefully 
and inspect it to ensure that it is not damaged 
during shipment. If damage is found, retain the 
original packing material and immediately notify 
the carrier and the relevant local Yokogawa 
Sales office. Make sure the Model Code and 
Serial Number on the sensor are the same as 
on the packing list.

1.3 Warranty and Service

Yokogawa products are guaranteed free from 
defects in workmanship and materials under 
normal use and service for a period of (typically) 
12 months from the date of shipment from the 
manufacturer. Individual sales organizations 
can deviate from the typical warranty period, 
and the conditions of sale relating to the 
original purchase order should be consulted. 
Damage caused by wear and tear, inadequate 
maintenance, corrosion, or by the effects of 
chemical processes is excluded from this 
warranty coverage. In the event of a warranty 
claim, the defective goods should be sent 
(freight paid) to the Service Department of the 
relevant Yokogawa Sales office for repair or 
replacement (at Yokogawa’s discretion). The 
following information must be included in the 
letter accompanying the returned goods:
•  Model Code and Serial Number.
•  Original Purchase Order and Date.
•  Length of time in service and description of 

the process.

•  Description of the fault and circumstances of 

the failure.

•  Process/environmental conditions that may 

be related to the failure of the sensor.

•  Statement as to whether warranty or non-

warranty service is requested.

IM12B06J07-02E-E_ed05.indd   6

14/01/16   09:46

Summary of Contents for SC24V

Page 1: ...IM 12B06J07 02E E 5th edition Instruction Manual Model SC24V Combined 12mm 4 in 1 differential pH sensor YOKOGAWA IM12B06J07 02E E_ed05 indd 1 14 01 16 09 46...

Page 2: ...les de instrucciones para los productos antiexplosivos de ATEX est n disponibles en ingl s Si desea solicitar las instrucciones de estos art culos antiexplosivos en su idioma local deber ponerse en co...

Page 3: ...xima ou com um representante da Yokogawa PL Wszystkie instrukcje obs ugi dla urz dze w wykonaniu przeciwwybuchowym Ex zgodnych z wymaganiami ATEX dost pne s w j zyku angielskim Je eli wymagana jest i...

Page 4: ...IM 12B06J07 02E E 4 IM12B06J07 02E E_ed05 indd 4 14 01 16 09 46...

Page 5: ...Shipping details 9 2 8 Environmental conditions 9 3 INSTALLATION OF SC24V 10 3 1 Typical installation 10 3 2 Preparing the sensor for use 10 3 3 Mounting the sensor 10 4 DIMENSIONS 13 5 WIRING 14 6 GE...

Page 6: ...ion The only applications where the SC24V cannot be used are those where the salt concentration is fluctuating more than 30 or is too low such as in pure and ultrapure water processes 1 2 Unpacking an...

Page 7: ...h the sensor options have been exposed 1 4 Serial Number definition The Serial Number is defined by nine 9 alphanumeric characters X1 X2 Production Location X3 X4 Year Month code X5 X6 X7 X8 X9 Tracki...

Page 8: ...4 Dynamic specifications Response time pH t90 15 sec for 7 to 4 pH step Response time temperature t90 1 5 min for 10 C step Stabilization time pH 2 min for 0 02 pH unit during 10 sec 2 5 Operating ra...

Page 9: ...e sensor has been connected to none intrinsically safe equipment which exceeds the restrictions regarding the sensor input circuit see electrical data the sensor is not suitable anymore for intrinsica...

Page 10: ...from its shipping box and slide off the so called wet pocket the tube filled with solution to prevent drying out of the measuring elements during shipment or storage During shipment electrolyte in the...

Page 11: ...ormation of these parts is given in Section 8 of this Instruction Manual For a detailed description of these adapters see Section 4 Examples of mounting the SC24V sensor using an adapter are given in...

Page 12: ...1 4 NPT Flow fitting SS K1598AC 1 4 NPT Adapter PG13 5 to SC fitting K1523JA Noryl K1523JC SS Slide ring O ring FF40 S22 Figure 6 Mounting of sensor in fitting K1598AC using adapter K1523JB K1523JD F...

Page 13: ...0 Figure 8 Dimensions adapters K1500DV K1520JN K1520JP Figure 9 Dimensions adapters K1523JA K1523JC Figure 10 Dimensions adapters K1523JB K1523JD Figure 7 Dimensions SC24V Sensor according to DIN 1926...

Page 14: ...ts Figure 12 WU10 dual coax cable A C D E F G H Indication groove B Green Yellow Black White White Core coax White Core coax Brown Brown Brown Red Blue 14 63 17 13 15 16 11 12 Table 4 Definition Vario...

Page 15: ...lyzer to MANUAL calibration Wait until the pH reading is stable Adjust the pH reading in the pH analyzer to the value indicated on the bottle in this case 7 00 Go to solution 2 to do a 2 point calibra...

Page 16: ...ave an open reference junction and is hermetically sealed from the process it does not suffer from poisoning diffusion and fouling This means the SC24V sensor requires very little maintenance In most...

Page 17: ...onic strength 1 mol NaCl K1520BK Buffer solution pH 9 pNa 0 3 x 500 ml ionic strength 1 mol NaCl Adapters K1500DV Adapter PG13 5 to M25x1 5 PVDF K1520JN Adapter PG13 5 to M25x1 5 PVC K1520JP Adapter P...

Page 18: ...nts EN 60079 11 2007 Explosive atmospheres Part 11 Equipment protection by intrinsic safety i EN 60079 26 2007 Explosive atmospheres Part 26 Equipment with equipment protection level EPL Ga The provis...

Page 19: ...Min ______________ Max ______________ Concentration of salts________________________Concentration of organics_____________________ Type s ____________________________________Type s ___________________...

Page 20: ...a com eu YOKOGAWA ELECTRIC CORPORATION World Headquarters 9 32 Nakacho 2 chome Musashino shi Tokyo 180 8750 Japan www yokogawa com YOKOGAWA CORPORATION OF AMERICA 2 Dart Road Newnan GA 30265 USA www y...

Reviews: