3  INSTALLATION OF SC24V 

For optimum measurement results, the SC24V 
should be installed in a location that offers 
an acceptable representation of the process 
composition and 

DOES NOT

 exceed the 

specifications of the sensor.
The SC24V is designed with PG13.5 threaded 
connection to allow installation in a wide variety 
of applications.

3.1 Typical installation

The SC24V sensor is designed for versatile 
in-line, immersion or off-line installation. For best 
results the SC24V should be mounted with the 
process flow coming towards the sensor, and 
positioned at least 15° above the horizontal 
plane to eliminate air bubbles in the pH glass 
bulb (see Figure 1).

Figure 1: 

Mounting positions of sensor

3.2 Preparing the sensor for use

Remove the sensor from its shipping box and 
slide off the so-called ‘wet pocket’, the tube 
filled with solution to prevent drying out of the 
measuring elements during shipment or storage. 
During shipment, electrolyte in the sensor could 
be dislocated. To correct this, place the sensor 
upright for 24 hours. Before installing the sensor 
in the process it should be calibrated. A general 

calibration procedure is described in Section 6 
of this Instruction Manual. 
 

3.3 Mounting the sensor

The simplest mounting is to use the PG13.5 
threaded connection of the sensor. The sensor 
is standard with a slide ring (Ryton) and an 
O-ring (Silicon) for direct mounting in a fitting 
provided with PG13.5 thread (see Figure 2). 

Other O-ring materials are available as a spare 
part (see Section 8 of this Instruction Manual).

Figure 2: 

Simple mounting of sensor in PR10 

retractable fitting

I N C O R R E C T

1 5 º

1 5 º

H O R I Z O N T A L

G O O D

G O O D

Slide ring

O-ring

SENSOR HOLDER

PG 13,5

IM12B06J07-02E-E_ed05.indd   10

14/01/16   09:47

Summary of Contents for SC24V

Page 1: ...IM 12B06J07 02E E 5th edition Instruction Manual Model SC24V Combined 12mm 4 in 1 differential pH sensor YOKOGAWA IM12B06J07 02E E_ed05 indd 1 14 01 16 09 46...

Page 2: ...les de instrucciones para los productos antiexplosivos de ATEX est n disponibles en ingl s Si desea solicitar las instrucciones de estos art culos antiexplosivos en su idioma local deber ponerse en co...

Page 3: ...xima ou com um representante da Yokogawa PL Wszystkie instrukcje obs ugi dla urz dze w wykonaniu przeciwwybuchowym Ex zgodnych z wymaganiami ATEX dost pne s w j zyku angielskim Je eli wymagana jest i...

Page 4: ...IM 12B06J07 02E E 4 IM12B06J07 02E E_ed05 indd 4 14 01 16 09 46...

Page 5: ...Shipping details 9 2 8 Environmental conditions 9 3 INSTALLATION OF SC24V 10 3 1 Typical installation 10 3 2 Preparing the sensor for use 10 3 3 Mounting the sensor 10 4 DIMENSIONS 13 5 WIRING 14 6 GE...

Page 6: ...ion The only applications where the SC24V cannot be used are those where the salt concentration is fluctuating more than 30 or is too low such as in pure and ultrapure water processes 1 2 Unpacking an...

Page 7: ...h the sensor options have been exposed 1 4 Serial Number definition The Serial Number is defined by nine 9 alphanumeric characters X1 X2 Production Location X3 X4 Year Month code X5 X6 X7 X8 X9 Tracki...

Page 8: ...4 Dynamic specifications Response time pH t90 15 sec for 7 to 4 pH step Response time temperature t90 1 5 min for 10 C step Stabilization time pH 2 min for 0 02 pH unit during 10 sec 2 5 Operating ra...

Page 9: ...e sensor has been connected to none intrinsically safe equipment which exceeds the restrictions regarding the sensor input circuit see electrical data the sensor is not suitable anymore for intrinsica...

Page 10: ...from its shipping box and slide off the so called wet pocket the tube filled with solution to prevent drying out of the measuring elements during shipment or storage During shipment electrolyte in the...

Page 11: ...ormation of these parts is given in Section 8 of this Instruction Manual For a detailed description of these adapters see Section 4 Examples of mounting the SC24V sensor using an adapter are given in...

Page 12: ...1 4 NPT Flow fitting SS K1598AC 1 4 NPT Adapter PG13 5 to SC fitting K1523JA Noryl K1523JC SS Slide ring O ring FF40 S22 Figure 6 Mounting of sensor in fitting K1598AC using adapter K1523JB K1523JD F...

Page 13: ...0 Figure 8 Dimensions adapters K1500DV K1520JN K1520JP Figure 9 Dimensions adapters K1523JA K1523JC Figure 10 Dimensions adapters K1523JB K1523JD Figure 7 Dimensions SC24V Sensor according to DIN 1926...

Page 14: ...ts Figure 12 WU10 dual coax cable A C D E F G H Indication groove B Green Yellow Black White White Core coax White Core coax Brown Brown Brown Red Blue 14 63 17 13 15 16 11 12 Table 4 Definition Vario...

Page 15: ...lyzer to MANUAL calibration Wait until the pH reading is stable Adjust the pH reading in the pH analyzer to the value indicated on the bottle in this case 7 00 Go to solution 2 to do a 2 point calibra...

Page 16: ...ave an open reference junction and is hermetically sealed from the process it does not suffer from poisoning diffusion and fouling This means the SC24V sensor requires very little maintenance In most...

Page 17: ...onic strength 1 mol NaCl K1520BK Buffer solution pH 9 pNa 0 3 x 500 ml ionic strength 1 mol NaCl Adapters K1500DV Adapter PG13 5 to M25x1 5 PVDF K1520JN Adapter PG13 5 to M25x1 5 PVC K1520JP Adapter P...

Page 18: ...nts EN 60079 11 2007 Explosive atmospheres Part 11 Equipment protection by intrinsic safety i EN 60079 26 2007 Explosive atmospheres Part 26 Equipment with equipment protection level EPL Ga The provis...

Page 19: ...Min ______________ Max ______________ Concentration of salts________________________Concentration of organics_____________________ Type s ____________________________________Type s ___________________...

Page 20: ...a com eu YOKOGAWA ELECTRIC CORPORATION World Headquarters 9 32 Nakacho 2 chome Musashino shi Tokyo 180 8750 Japan www yokogawa com YOKOGAWA CORPORATION OF AMERICA 2 Dart Road Newnan GA 30265 USA www y...

Reviews: