background image

IM 12B06J07-02E-E

15 

Calibration of the SC24V sensor has to be done 
with the pH analyzer connected. Normally the 
pH standards that are preprogrammed in the 
pH analyzer can be used for calibration with the 
pH analyzer set to 

“AUTOCAL”

. The SC24V 

however is a differential pH sensor which needs 
pH buffers that have the same ionic strength 
because the sodium reference will change as 
the ionic strength changes. These pH buffers 
are not preprogrammed in the pH analyzer, 
which means the calibration has to be done 
with specific buffer solutions (see Section 
6.1) and the pH analyzer set to

 “MANUAL”

 

calibration. Refer to the analyzer Instruction 
Manual for details. Calibration data of the 
sensor is stored into the pH analyzer.

6.1 Buffer calibration

To calibrate the SC24V sensor, two buffer 
solutions with known pH values are required. 
It is recommended that one buffer solution has 
a value near to pH 7.00. Depending on the 
process value to be measured, the second 
buffer solution should be either acidic  
(below pH 7.00) or alkaline (above pH 7.00). 
Buffers which are available are: pH 4.00,  
pH 7.00 and pH 9.00. 
See Section 8 for ordering information.

The following is a very general 2-point manual 
calibration procedure with buffer solution:
1. Clean the sensor using a 5% solution of HCl;
2. Rinse sensor thoroughly with tap water  

(

DO NOT 

use demineralized water);

3. Immerse the sensor in the first buffer solution 

(pH 7.00 is recommended); 
Set the pH analyzer to 

“MANUAL”

 

calibration. Wait until the pH reading is stable. 
Adjust the pH reading in the pH analyzer 
to the value indicated on the bottle (in this 

case 7.00). Go to solution 2 to do a 2-point 
calibration;

4. Rinse sensor thoroughly with tap water;
5. Immerse the sensor in the second buffer  

(pH 4.00 is recommended); 
Wait until the pH reading is stable. Adjust the 
pH reading in the pH analyzer to the value 
indicated on the bottle (in this case 4.00).  
Quit the manual calibration routine by 
accepting the new calculated calibration data 
of zero and slope.

After calibration, re-install the sensor into the 
process.

Note:

  It is important to understand it is possible 

the SC24V differential sensor does 
not show the correct pH value after 
buffer calibration. The reason is that the 
ionic strength of the buffers is 1 mol 
NaCl, which can deviate from the ionic 
strength of the process. For an accurate 
pH reading an extra 1 point process 
calibration has to be done (see Section 
6.2) to compensate for differences in 
ionic strength. 

6.2 Process calibration

The following is a specific 1-point manual 
calibration procedure for the process to 
measure:
Set the pH analyzer to “

MANUAL

” calibration. 

Wait until the pH reading is stable. 
Adjust the pH reading in the pH analyzer to the 
value of the process sample that is analyzed 
using a laboratory measurement. For the 
laboratory measurement, we advice to use the 
Yokogawa PH72 personal pH meter. 
Quit the manual calibration routine by accepting 
the new calculated calibration data of zero. 
Slope value will be unchanged.

6  GENERAL CALIBRATION & MAINTENANCE PROCEDURE

IM12B06J07-02E-E_ed05.indd   15

14/01/16   09:47

Summary of Contents for SC24V

Page 1: ...IM 12B06J07 02E E 5th edition Instruction Manual Model SC24V Combined 12mm 4 in 1 differential pH sensor YOKOGAWA IM12B06J07 02E E_ed05 indd 1 14 01 16 09 46...

Page 2: ...les de instrucciones para los productos antiexplosivos de ATEX est n disponibles en ingl s Si desea solicitar las instrucciones de estos art culos antiexplosivos en su idioma local deber ponerse en co...

Page 3: ...xima ou com um representante da Yokogawa PL Wszystkie instrukcje obs ugi dla urz dze w wykonaniu przeciwwybuchowym Ex zgodnych z wymaganiami ATEX dost pne s w j zyku angielskim Je eli wymagana jest i...

Page 4: ...IM 12B06J07 02E E 4 IM12B06J07 02E E_ed05 indd 4 14 01 16 09 46...

Page 5: ...Shipping details 9 2 8 Environmental conditions 9 3 INSTALLATION OF SC24V 10 3 1 Typical installation 10 3 2 Preparing the sensor for use 10 3 3 Mounting the sensor 10 4 DIMENSIONS 13 5 WIRING 14 6 GE...

Page 6: ...ion The only applications where the SC24V cannot be used are those where the salt concentration is fluctuating more than 30 or is too low such as in pure and ultrapure water processes 1 2 Unpacking an...

Page 7: ...h the sensor options have been exposed 1 4 Serial Number definition The Serial Number is defined by nine 9 alphanumeric characters X1 X2 Production Location X3 X4 Year Month code X5 X6 X7 X8 X9 Tracki...

Page 8: ...4 Dynamic specifications Response time pH t90 15 sec for 7 to 4 pH step Response time temperature t90 1 5 min for 10 C step Stabilization time pH 2 min for 0 02 pH unit during 10 sec 2 5 Operating ra...

Page 9: ...e sensor has been connected to none intrinsically safe equipment which exceeds the restrictions regarding the sensor input circuit see electrical data the sensor is not suitable anymore for intrinsica...

Page 10: ...from its shipping box and slide off the so called wet pocket the tube filled with solution to prevent drying out of the measuring elements during shipment or storage During shipment electrolyte in the...

Page 11: ...ormation of these parts is given in Section 8 of this Instruction Manual For a detailed description of these adapters see Section 4 Examples of mounting the SC24V sensor using an adapter are given in...

Page 12: ...1 4 NPT Flow fitting SS K1598AC 1 4 NPT Adapter PG13 5 to SC fitting K1523JA Noryl K1523JC SS Slide ring O ring FF40 S22 Figure 6 Mounting of sensor in fitting K1598AC using adapter K1523JB K1523JD F...

Page 13: ...0 Figure 8 Dimensions adapters K1500DV K1520JN K1520JP Figure 9 Dimensions adapters K1523JA K1523JC Figure 10 Dimensions adapters K1523JB K1523JD Figure 7 Dimensions SC24V Sensor according to DIN 1926...

Page 14: ...ts Figure 12 WU10 dual coax cable A C D E F G H Indication groove B Green Yellow Black White White Core coax White Core coax Brown Brown Brown Red Blue 14 63 17 13 15 16 11 12 Table 4 Definition Vario...

Page 15: ...lyzer to MANUAL calibration Wait until the pH reading is stable Adjust the pH reading in the pH analyzer to the value indicated on the bottle in this case 7 00 Go to solution 2 to do a 2 point calibra...

Page 16: ...ave an open reference junction and is hermetically sealed from the process it does not suffer from poisoning diffusion and fouling This means the SC24V sensor requires very little maintenance In most...

Page 17: ...onic strength 1 mol NaCl K1520BK Buffer solution pH 9 pNa 0 3 x 500 ml ionic strength 1 mol NaCl Adapters K1500DV Adapter PG13 5 to M25x1 5 PVDF K1520JN Adapter PG13 5 to M25x1 5 PVC K1520JP Adapter P...

Page 18: ...nts EN 60079 11 2007 Explosive atmospheres Part 11 Equipment protection by intrinsic safety i EN 60079 26 2007 Explosive atmospheres Part 26 Equipment with equipment protection level EPL Ga The provis...

Page 19: ...Min ______________ Max ______________ Concentration of salts________________________Concentration of organics_____________________ Type s ____________________________________Type s ___________________...

Page 20: ...a com eu YOKOGAWA ELECTRIC CORPORATION World Headquarters 9 32 Nakacho 2 chome Musashino shi Tokyo 180 8750 Japan www yokogawa com YOKOGAWA CORPORATION OF AMERICA 2 Dart Road Newnan GA 30265 USA www y...

Reviews: