< 2. Installation, Piping and Wiring >
2-4
IM 12B07N01-01E
(4) When a sensor with medium pressure KCl, solution tank is used:
Provide stop valves in the up-line and down-line piping adjacent to the holder.
Stop Valve
Stop Valve
F2.7E.ai
Figure 2.7
Stop Valve Installation
2.2.2
Piping Procedure
Piping Materials:
Use the materials shown below for the process piping adjacent to the flow-through type holder.
• Hard polyvinylchloride pipe
Nominal diameter 25 mm
• Polypropylene pipe
Nominal diameter 25 mm
• Wire rainforced soft polyvinylchloride pipe
Nominal diameter 25 mm
• Stainless steel pipe (JIS G3459)
SUS304 or SUS316
Nominal diameter 25 mm
Piping Example:
1. The outlet tube should be as short as possible to expel
the solution to atmosphere.
2. When a horizontal outlet tubing is long,
a vertical tube (H) should also be long.
Flowmeter
Flowmeter
F2.8E.ai
H
Figure 2.8
Sensor with General Type Reserve Tank