< 1. Specifications >
1-13
IM 12B07N01-01E
5.
Solenoid Valve
UNIT: mm
General Purpose Solenoid Valve
PH8MV*D
Explosionproof Solenoid Valve
PH8MVF*B
F1.5.ai
Cautions on Installation of Solenoid Valve for Jet / Brush Cleaning
1. Do not allow a sample solution to flow backward into the solenoid valve or to be replaced with the driving fluid.
For this take relevant measures; e.g. install a check valve to prevent inverse pressure between the inlet and outlet of the solenoid
valve, or install the solenoid valve higher than the holder, especially when using the air jet/brush cleaning system.
2. Make sure to avoid the risk of corrosion of the solenoid body (bronze) and seal (nitrilel rubber) by vapor or gaseous components
generated from a sample solution, especially when using the air jet/brush cleaning system.
G1/2
2-Rc1/2
G1/2
2-Rc1/2
103
102
71
14.5
46
72
95
70
90
146.5
64
14
Figure 1.5
Solenoid Valve for Jet & Brush Cleaning