IM 12B6C2-E-H
39
Class I, Division I, Groups ABCD
Vmax
= 31.5 V
Imax
= 93.3 mA
Ci
= 5 µF
Li
= 16.5 mH
8-2. FM
A)
1 ORP-sensor,
such as (but not limited to)
SM29-PT8
or
A)
1 PH-galss sensor,
SM21-AG4
such as (but not limited to)
SM21-AG6
plus
SM21-TL4
B)
1 reference sensor
SR20-AC22
such as (but not limited to)
SR20-AP24
or
SR20-AL32
AB) 1 combined sensor
SC21-AGP24
such as (but not limited to)
SC21-ASP24
plus
C)
1 temperature sensor
SM60-Pt100
such as (but not limited to)
SM60-Pt1000
These are all passive sensors to be regarded as “simple appara-
tus”, i.e. devices which do neither store nor generate voltages
over 1.2 V, currents over 0.1 A, power over 25 W or energy over
20 mJ .
pH 200 S
Vout
= 14.4 V max.
Iout
= 26 mA max.
C ext
= 1 µF max
L ext
= 58 mH max.
SC 200 S
Vout
= 14.4 V max.
Iout
= 129 mA max.
C ext
= 1 µF max
L ext
= 2.3 mH max.
NON HAZARDOUS LOCATION
31.5 Vmax
4-20 mA
CONTROL DRAWING FOR EXA 200 INSTRUMENTS
HAZARDOUS LOCATION
1
conductivity sensor with
SC41-SP34
integral temperature sensor
SC49-FP08
such as (but not limited to)
SC41-SP24
SC41-EP14
SC41-EP04
These are all passive sensors to be regarded as “simple appara-
tus”, i.e. devices which do neither store nor generate voltages
over 1.2 V, currents over 0.1 A, power over 25 W or energy over
20 mJ .
Connecting cables
such as (but not limited to):
WU20-PC5
WU20-PC2
WU20-PC10
K1220 EM
WU-LT2
10 meters maximum
Connecting cables
such as (but not limited to):
WU40-LH18
WU40-LH2
WU40-LH5
10 meters maximum
FMRC-approved barrier
To be installed in accordance with ANSI/RP 12.6 and NEC requirements
Maximum safe area voltage should not exceed 250 Vrms
+
+
–
–
•
•
•
11 12 14 13 17 15 16
11 12 14 13 17 15 16