4-3. Set up the HOLD-function
IM 12B6C2-E-H
1. Access to HOLD-function
Remove cover by releasing 4 screws.
Push this key to select
Commissioning mode
Select HOLD
Move pointer to SET HOLD
HOLD = HOLD-function
Confirm selection
2. Activate HOLD-function
Display shows actual status
*H.OFF
= HOLD not active
*H.ON
= HOLD activated
Activate (deactivate) HOLD
Changing setting
Confirm setting
*H.FIX
= HOLD fixed value
*H.LST
= HOLD last value
Select HOLD fixed or last
Change selection
Confirm selection
3. Select value to hold
Display shows current setting
Adjust fixed value
Select digit to adjust
Adjust value of digit
Confirm setting
Return to measurement
14
OUTPUT
SET HOLD
SERVICE
YES
NO
YES
NO
MODE
*
.
OUTPUT
SET HOLD
SERVICE
YES
NO
MODE
ENT
.
YES
ENT
>
^
NO
YES
NO
YES
NO
x
ESCAPE TO MEASURE can be used at any stage to abort operation.
WARNING: If the HOLD function is activated the instrument returns with the question HOLD (flashing); answer YES or NO or
MODE again to return to measurement.
MODE
MODE
OUTPUT
SET HOLD
SERVICE
YES
NO
YES
NO
MODE
*