5-2. Manual calibration
IM 12B6C2-E-H
1. Access to manual calibration
MAN.CAL = Manual calibration
Access to maintenance mode
Select manual calibration
Move pointer to MAN.CAL
Confirm selection
2. Adjust value manually
Display shows measured value
START = Start manual calibration
Confirm start of adjustment (dis-
play frozen, ready for adjustment
to new value)
Adjust new value.
Adjust process value to previously determi-
ned value (e.g. from hand-held pH-meter)
Select digit to adjust
Adjust value
Confirm adjusted value
3. End calibration
CAL.END = End of calibration
End calibration
End calibration and return to
measurement
Continue routine to calibrate
second point when using buffer
solution (repeat step 2)
Error E02 or E19 indicate manual calibration
is not acceptable